590625
39
Zoom out
Zoom in
Previous page
1/74
Next page
GMC-I Messtechnik GmbH 39
V/Hz/dB – Ω, Temperature, and A/Hz Measurements
5.5 Temperature Measurement
Temperature measurement is performed with a type K thermocouple or a
type Pt100 or Pt1000 resistance sensor (accessory, not included) which
is connected to the voltage input.
Selecting the Unit of Measure for Temperature (°C = default value)
1nFo
... set t iME ... tempunit
unit set: °C / °F

5.5.1 Measurement with Thermocouples, Temp TC
Ð Set the rotary switch to “Temp
TC
”.
Note
The last selected temperature measurement or the last selected
temperature sensor, i.e. type K or Pt100/Pt1000, remains in
memory and is accordingly displayed. Press the
FUNC | ENTER
key in order to change to the other measuring function.
Ð Either the internal reference junction (“1tEMPintern” parameter,
see also page 52) or an external reference temperature can be
specified as the reference temperature (see page 52). The type
(“I.” for internal or “E.” for external) and temperature of the se-
lected reference junction appears at the right-hand auxiliary
display during measurement.
Note
The internal reference temperature (temperature of the inter-
nal reference junction) is measured by a temperature sensor
inside of the instrument. This may deviate somewhat from
room temperature as a result of internal heat-up, or moving
from warmer to colder surroundings or vice versa.
Ð Connect the sensor to the two accessible jacks. The instru-
ment displays the measured temperature using the selected
unit of measure.
Note
After previously performing a 10 A current measurement, the
measuring instrument should be allowed to cool down for
30 minutes before performing measurements with thermo-
couples, in order to assure that the specified accuracy levels
are achieved.
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
Temp
TC
0023
.
3
°C
Temp
RTD
TC
TC
0256.0
°C
Pt1000
RTD
Measuring Range
TC K (NiCr-Ni) 270.0 +1372.0 °C
FUNC
ENTER
1
.
22
.
4
COM
V
Tem p
Ω
TC
39


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRAHIT ULTRA BT at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRAHIT ULTRA BT in the language / languages: English as an attachment in your email.

The manual is 0,9 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info