16 GMC-I Messtechnik GmbH
Note!
Testing the Measurement Cables
Initial testing should be performed by short-circuiting
the measurement cables using their test probes. The
instrument must display nearly 0 Ω. A broken
measurement cable can be detected in this way.
Attention!
!
Do not touch the conductive ends of the test probes
when the instrument has been activated for the
measurement of insulation resistance.
You may otherwise be exposed to a current of
1.5 mA (limited in the measuring instrument), and
although this is not life endangering, the resulting
electrical shock is quite discernible.
If, on the other hand, measurement is being performed on
a capacitive device under test, for example a cable, it may
be charged to approximately ±600 V, depending upon the
selected nominal voltage.
Touching the device under test after measurement has
been performed may be life endangering in this case!
11.2 Insulation Resistance Measurement
Ð Press and hold the FUNC key until the display settles in, in
order to measure insulation resistance.
Attention!
!
The selected and displayed test voltage blinks during
measurement.
Attention: Contact hazard!
Current voltage at the device under test appears at the
interference or residual voltage display during measurement.
This value is slightly less than nominal voltage.
Insulation resistance measurement is ended by releasing the
FUNC key.
Auto-ranging is active during insulation resistance
measurement. Manual measuring range selection is not
possible.
Note!
The instrument’s storage batteries are rapidly
depleted during insulation resistance measurement.
Only hold the multifunction key depressed as long as
is actually necessary. The continuous measurement
described below should only be performed if
absolutely necessary.
Use NiMH storage batteries only.
Continuous Measurement
Ð Activation: Briefly press the multifunction key and the
MAN|AUTO key simultaneously. This is acknowledged with
an acoustic signal.
Ð Deactivation: Briefly press the multifunction key.
11.3 Ending the Measurement and Discharging
After measurement has been completed, any remaining
residual voltage is displayed which may result from cable
capacitance. These charges are quickly dispersed by means
of two internal resistors with a value of 2 MΩ. However,
contact to the device under test must be maintained. The
falling voltage value can be observed at the interference or
residual voltage display.
Do not disconnect the device under test until the voltage value has
dropped to below 25 V!
mΩ
S+S–
MΩ
R
ISO
U
Measured
(50 to 500V)
Ω: 10 k
Ω 3GΩ
Measuring Range:
!
16


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Gossen Metrawatt METRA HIT 27I at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Gossen Metrawatt METRA HIT 27I in the language / languages: English as an attachment in your email.

The manual is 0,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Gossen Metrawatt METRA HIT 27I

Gossen Metrawatt METRA HIT 27I Quick start guide - English, German - 4 pages

Gossen Metrawatt METRA HIT 27I User Manual - Dutch - 28 pages

Gossen Metrawatt METRA HIT 27I Quick start guide - French - 4 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info