8845A/8846A
Users Manual
1-22
Capacitance (8846A Only)
Accuracy is stated as ±(% of measurement + % of range)
Range Resolution
1 Year Accuracy [1]
(23 ±5 °C)
Temperature Coefficient/ °C
Outside 18 to 28 °C
1 nF 1 pF 2% ± 2.5 % 0.05 + 0.05
10 nF 10 pF 1% ± 0.5 % 0.05 + 0.01
100 nF 100 pF 1% ± 0.5 % 0.01 + 0.01
1 μF 1 nF 1% ± 0.5 % 0.01 + 0.01
10 μF 10 nF 1% ± 0.5 % 0.01 + 0.01
100 μF 100 nF 1% ± 0.5 % 0.01 + 0.01
1 mF 1 μF 1% ± 0.5 % 0.01 + 0.01
10 mF 10 μF 1% ± 0.5 % 0.01 + 0.01
100 mF 100 μF 4% ± 0.2 % 0.05 + 0.05
[1] Stated accuracy is attained when Zero function is used.
Temperature (8846A only)
Test Current ............................................................ 1 mA
Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 ohms lead
resistance. The accuracy listed in the table below are valid only when using the 4-wire RTD measurement function.
Specifications do not include probe accuracy, which must be added.
Accuracy
Range Resolution
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
-200 °C 0.001 °C 0.06 0.09 0.0025
-100 °C 0.001 °C 0.05 0.08 0.002
0 °C 0.001 °C 0.04 0.06 0.002
100 °C 0.001 °C 0.05 0.08 0.002
300 °C 0.001 °C 0.1 0.12 0.002
600 °C 0.001 °C 0.18 0.22 0.002
Additional Errors
Digits NPLC Additional NPLC Noise Error
6 ½ 100 0 °C
6 ½ 10 0 °C
5 ½ 1 0.03 °C
5 ½ 0.2 0.12 °C
4 ½ 0.02 0.6 °C
Continuity
Continuity Threshold ............................................... Selectable between 1 Ω and 1000 Ω
Test Current ............................................................ 1 mA
Response Time....................................................... 300 samples/sec with audible tone
Accuracy is given as ± (% measurements + % of range)
Range 24 Hour
(23 ±1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
1000.0 Ω 0.002 + 0.01 0.008 + 0.02 0.01 + 0.02 0.001 + 0.002
32


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8845A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8845A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8845A

Fluke 8845A User Manual - German - 108 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info