8845A/8846A
Users Manual
1-20
8845A Accuracy
Accuracy is given as ± (% measurement + % of range)
Range Frequency
(Hz)
24 Hour
(23 ±1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
3 – 5 Hz 1.1 + 0.06 1.1 + 0.06 1.1 + 0.06 0.2 + 0.006
5 – 10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.1 + 0.006
10 Hz – 5 kHz 0.15 + 0.06 0.15 + 0.06 0.15+ 0.06 0.015 + 0.006
10 mA
5 – 10 kHz 0.35 + 0.7 0.35 + 0.7 0.35 + 0.7 0.03 + 0.006
3 – 5 Hz 1.0 + 0.04 1.0 + 0.04 1.0 + 0.04 0.1 + 0.006
5 – 10 Hz 0.3 + 0.04 0.3 + 0.04 0.3 + 0.04 0.035 + 0.006
10 Hz – 5 kHz 0.1 + 0.04 0.1 + 0.04 0.1 + 0.04 0.015 + 0.006
100 mA
5 – 10 kHz 0.2 + 0.25 0.2 + 0.25 0.2 + 0.25 0.03 + 0.006
3 – 5 Hz 1.0 + 0.1 1.0 + 0.1 1.0 + 0.1 0.1 + 0.006
5 – 10 Hz 0.3 + 0.1 0.3 + 0.1 0.3 + 0.1 0.035 + 0.006
10 Hz – 1 kHz 0.1 + 0.1 0.1 + 0.1 0.1 + 0.1 0.015 + 0.006
400 mA [3]
1kHz – 10 kHz 0.2 + 0.7 0.2 + 0.7 0.2 + 0.7 0.03 + 0.006
3 – 5 Hz 1.0 + 0.04 1.0 + 0.04 1.0 + 0.04 0.1 + 0.006
5 – 10 Hz 0.3 + 0.04 0.3 + 0.04 0.3 + 0.04 0.035 + 0.006
10 Hz – 5 kHz 0.1 + 0.04 0.1 + 0.04 0.1 + 0.04 0.015 + 0.006
1 A [2]
5 – 10 kHz 0.35 + 0.7 0.35 + 0.7 0.35 + 0.7 0.03 + 0.006
3 – 5 Hz 1.1 + 0.06 1.1 + 0.06 1.1 + 0.06 0.1 + 0.006
5 – 10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.035 + 0.006
10 Hz – 5 kHz 0.15 + 0.06 0.15 + 0.06 0.15 + 0.06 0.015 + 0.006
3 A [1][2]
5 – 10 kHz 0.35 + 0.7 0.35 + 0.7 0.35 + 0.7 0.03 + 0.006
3 – 5 Hz 1.1 + 0.06 1.1 + 0.06 1.1 + 0.06 0.2 + 0.006
5 – 10 Hz 0.35 + 0.06 0.35 + 0.06 0.35 + 0.06 0.035 + 0.006
10 Hz – 5 kHz 0.15 + 0.06 0.15 + 0.06 0.15 + 0.06 0.015 + 0.006
10 A [2]
5 – 10 kHz 0.35 + 0.7 0.35 + 0.7 0.35 + 0.7 0.03 + 0.006
[1] Part of the 10 A range
[2] Available only at front panel connectors
[3] 400 mA available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on,
1 minute off; maximum crest factor 3:1 at 400 mA; specification for current above 329 mA is typical.
Additional Low Frequency Errors
Error is stated as % of reading.
AC Filter
Frequency 3HZ (slow) 20HZ (medium) 200HZ (fast)
10 – 20 Hz 0 0.25
20 – 40 Hz 0 0.02
40 – 100 Hz 0 0.01 0.55
100 – 200 Hz 0 0 0.2
200 Hz – 1 kHz 0 0 0.02
> 1 kHz 0 0 0
Frequency
Gate Times ............................................................. Programmable to 1 s, 100 ms, and 10 ms
Measurement Method ............................................. Flexible counting technique. AC-coupled input using the ac voltage
measurement function.
Settling Considerations........................................... When measuring frequency or period after a dc offset voltage change,
errors may occur. For the most accurate measurement, wait up to
1 second for the input blocking capacitor to settle.
Measurement Considerations ................................. To minimize measurement errors, shield inputs from external noise
when measuring low-voltage, low-frequency signals.
30


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8845A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8845A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8845A

Fluke 8845A User Manual - German - 108 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info