763941
33
Zoom out
Zoom in
Previous page
1/104
Next page
Introduction and Specifications
Electrical Specifications 1
1-23
Diode Test
Test Current ............................................................ 100 μA or 1 mA
Response Time....................................................... 300 samples/sec with audible tone.
Accuracy is given as ± (% measurements + % of range)
Range 24 Hour
(23 ±1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
Temperature
Coefficient/ °C
Outside 18 to 28 °C
5.0000 V 0.002 + 0.002 0.008 + 0.002 0.01 + 0.002 0.001 + 0.002
10.0000 V 0.002 + 0.001 0.008 + 0.002 0.01 + 0.002 0.001 + 0.002
Measurement Rates (IEEE488[4])
Measurements/Second [1]
Function Digits Setting
Integration Time
60 Hz (50 Hz) 8845A 8846A
100 NPLC 1.67 (2) s 0.6 (0.5) 0.6 (0.5)
10 NPLC 167 (200) ms 6 (5) 6 (5)
1 NPLC 16.7 (20) ms 60 (50) 60 (50)
0.2 NPLC 3.3 ms 270 270
DC Volts, DC Current, and
Resistance
0.02 NPLC 500 us 995 995
6½ 3 Hz 0.47 0.47
6½ 20 Hz 1.64 1.64
AC Voltage and AC Current [2]
6½ 200 Hz
[3] 4.5 4.5
6½ 1 s 1 1
5½ 100 ms 9.8 9.8
Frequency and Period
4½ 10 ms 80 80
Capacitance 6½ NA 2
[1] Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off.
[2] Maximum measurement rates for 0.01 % of ac step. When dc input varies, additional settling delay is required.
[3] For remote operation or external trigger using default settling delay
[4] Speeds available in OutG SW 1.0.700.18 or higher. Note that the measurements rates for RS232 can vary depending on the
baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurement/s. The LAN bus
has a maximum measurement rate of 963 measurement/s.
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8845A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8845A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8845A

Fluke 8845A User Manual - German - 108 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info