Operating the Meter Using the Computer Interface
Computer Interface Command Set 4
4-15
Table 4-7. Description of Bits in the Status Byte Register (STB) (cont.)
Bit No. Name Condition
6 Master Summary Status (MSS) [1] Set to 1 if any enabled bit in the STB (MSS) register is set
to 1; otherwise set to 0. To determine the status of MSS
bit, send STB? query command.
Request Service (RQS) is set to 1 if service requested
from front panel or MSS is set to 1. Status of bit is
returned by serial poll, which clears RQS.
7 Not used Always set to 0.
[1] As read by *STB? Command. If the STB is read by a serial poll, bit 6 is returned as RQS.
Reading the Status Byte Register
The host reads the STB by taking a serial poll or sending the Meter a *STB? query. (The
value of the status byte is not affected by the STB? query.) When the STB is read, an
integer is returned. This integer is the decimal equivalent of an 8-bit binary number. For
example, 48 is the decimal equivalent of the binary 00110000, which means that bit 4
(MAV) and bit 5 (ESB) are set to 1.
If the status byte is read with an *STB? query, bit 6 is returned as Master Summary
Status (MSS).
See the following example:
*STB? reads the STB. If 32 is returned, it is converted to its binary equivalent of
00100000, which indicates that bit 5 (ESB) is set to 1. To determine the event status,
you would read the ESB in the same manner, using the *ESR? command.
Computer Interface Command Set
The remainder of this chapter describes the RS-232 computer interface commands.
Commands are grouped by related function and are listed in the tables that follow.
Parameters that must be supplied by the user or strings returned by the Meter are
enclosed in angle brackets (for example, <value>).
75


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8808A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8808A in the language / languages: English as an attachment in your email.

The manual is 1,96 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8808A

Fluke 8808A User Manual - German - 102 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info