763943
126
Zoom out
Zoom in
Previous page
1/134
Next page
8588A/8558A
Operators Manual
122
The sequence is:
At midnight on 5th October 2018 the trigger process leaves the Initiate layer and
passes (without a delay) through the ARM2 layer which is set at its defaults. As
the ARM1 source is Timer, the process continues down through the Arm1 event
detector starting the Arm1 Timer and incrementing the Arm1 layer loop counter
on the way.
The Trigger layer Timer starts as the process passes through the Trigger layer
Event detector. The trigger layer Loop counter is incremented and the first of the
reading acquisitions is triggered. The Trigger layer loop count (1) is less than
Count (50) so the upward process flow is diverted back through to the Trigger
layer Event detector. The process pauses at this point until the 30 second Timer
expires and another reading is triggered. This loop repeats until the Trigger layer
Loop counter equals Count (50) at which point the process flows up to the Arm1
layer.
Arm1 layer Loop Counter (1) equals Arm1 layer Count so the flow continues
upwards through Arm2 back to the initiate layer. The process so far has taken 50
x 30 = 1500 seconds = 25 minutes so the time is 2018:10:05:00:25:00. This is
less than the Epoch expiry so the process goes back down through Arm2 and
waits at the Arm1 Event detector until Arm1 Timer (3600 seconds) expires 35
minutes later at 2018:10:05:01:00:00. The process enters the Trigger layer to
begin the second group of 50 readings and the 3 day process continues.
At 2018:10:07:23:00:00 the Arm1 Timer expires at the start of the final circuit
around the Trigger subsystem. 25 minutes later the last group of 50 readings has
been gathered and the process flow returns to the Initiate layer. This time the
Epoch has expired so the system stops taking readings.
126


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8588A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8588A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8588A

Fluke 8588A User Manual - German - 134 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info