Reference Multimeter and 8 ½ Digit Multimeter
Triggering Measurements
121
Note
An alternative method to see the average of 5000 readings is to use
the Math function to set a Block Average of 5000 readings. With
Block averaging set, the display shows one reading for every trigger,
the value being the average of 5000 individual 200
μ
s
measurements, representing the topline of the applied square wave.
As usual, can be used to make this a continuous
measurement. If continuous were the desired behavior originally, the
Trigger/Arms Count is irrelevant in this particular example and could
have been left at 1. The MATH is performing the count.
Example 8
Measurement: Starting at midnight on 5th October 2018 you want the Product to
take a group of 50 readings spaced at 30 second intervals starting on the hour
for 3 days. The Product is in the default trigger mode (free-run) and remote
commands are used instead of the front panel.
Solution: Use the Epoch mode to set the start and stop times. When Epoch is
the selected Initiate mode, Initiate is Continuous OFF outside the Epoch times,
Continuous ON between them.
The Trigger layer Timer is set to 30 seconds to determine the spacing of the
individual readings. Starting from the default state the bus commands are:
TRIGger:SOURce TIMer
TRIGger:TIMer 30
TRIGger:DELay 0
The Trigger layer Count is set so 50 readings are taken before exit upwards from
the Trigger layer into the Arm1 layer.
TRIGger:COUNt 50
Arm1 layer Timer is set to 3600 seconds (1 hour)
ARM1:TIMer 3600
The Epoch start is set to the start time of midnight on 5th October 2018. The end
time is set just after the last hourly sequence starts.
INIT:EPOCh 2018:10:05:00:00:00, 2018:10:07:23:01:00
125


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8588A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8588A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8588A

Fluke 8588A User Manual - German - 134 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info