8588A/8558A
Operators Manual
24
In auto, the gate time is the longest of the cardinal times that will not reduce the
reading rate. If the gate time is set manually, the reading rate is the longest of
RMS filter and gate times.
ACV measurements wait for both the RMS filter settling and the counter gate,
whichever is the longer. If you select long counter gate times, this may slow the
reading rate. The auto times are chosen to not slow the read rate.
Peak to peak method: This submenu is active when the Secondary Reading
is set to Pk to Pk. Measured shows the peak to peak as measured in ACV
assuming no particular signal wave form. Sine, Square, Triangle, and
Truncated Sine specify the signal wave form that is measured, and calculates
the peak to peak based on the rms value. For example, if set to Sine, the
peak to peak displayed is 2 x (square root of 2) x rms. Square is 2 x rms,
Triangle is 2 x (square root of 3) x rms, and Truncated Sine is 4.618803 x
rms. Use the Square, Triangle and Truncated Sine selections to measure the
peak-to-peak output of multi-product calibrators like the Fluke 5522A which
have these non-sine wave outputs.
Measure AC Voltage
The sections below explain how to accurately measure ac voltage.
Induced Interference
If interference signals are present or lead interference (noise) takes place during
ac measurements, any induced interfering signals combine with the measured
signal to result in measurement errors. In some circumstances, it may be
possible to filter out the unwanted external signals, but it is generally more
effective to reduce the interference before it is induced. Accomplish this by
operating in a quiet environment, for example, using a screened cage if possible
and using twisted or shielded measurement leads as discussed below.
Common Mode Rejection
The principles of external guarding, outlined in the description of dc voltage
measurement, apply generally to ac voltage measurement. For ac, you gain
further advantage by using the external guard as a shield for the input leads.
Lead Considerations
In all cases, improve ac voltage measurement accuracy by shortening the leads
to the minimum-practical length. Doing this reduces lead capacitance, lead
inductance, and loop area.
Fluke Calibration recommends shielded twisted pair leads for low-frequency
measurements and coaxial leads for low and high frequency measurements.
Take care to avoid measurement errors from the interaction of lead capacitance
and inductance with any source output impedance. For additional information
and guidance, see the Fluke publication Calibration: Philosophy in Practice
(ISBN 0-9638650-0-5). See the ACV Reading Rate for RMS Filter Settings
Specification in the specifications. See Specifications.
28


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8558A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8558A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8558A

Fluke 8558A User Manual - German - 134 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info