763944
26
Zoom out
Zoom in
Previous page
1/134
Next page
8588A/8558A
Operators Manual
22
(RMS Filter): Push to select various filters for the rms converter, allowing
measurements down to the chosen filter frequency without degradation of
accuracy and excessive reading variation. One of the filters is always in circuit.
The 40 Hz filter is the default selection at power on. The filter choices available
are 0.1 Hz, 1 Hz, 10 Hz, 40 Hz, 100 Hz and 1 kHz. See Specifications. Use the
softkeys or the navigational keys to highlight the selection and then push .
Push to return the Product to the previous menu.
(Measure Setup): Has parameters that can be set up for ac voltage
measurements. Use the softkeys or the navigational keys to highlight the
selection and then push . Push to return the Product to the
previous menu. See the screen below:
iei022.png
The parameters under this menu are:
Signal path coupling, impedance: (note that this selection determines what
is available in Frequency path coupling) these different combinations of signal
path coupling and impedance are available:
o (AC, 1 MΩ)
o (DC, 1 MΩ)
o (AC, 10 MΩ)
o (DC, 10MΩ)
o (DC, Auto)
Most applications should use the 1 MΩ input impedance (default) as the
10 MΩ input has relaxed specifications. DC Auto selects the highest
impedance available for any given range.
26


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8558A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8558A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8558A

Fluke 8558A User Manual - German - 134 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info