8588A/8558A
Operators Manual
20
Common Mode Rejection - Use of External Guard Connection
When the source presents an unbalanced impedance to the measuring
terminals, and common mode voltages are present, use the GUARD terminal
with External Guard selected. Use and then (Ext. Guard) to
activate the GUARD terminal. See Input Terminal Selection (INPUTS).
Regardless of how the INPUT HI and LO terminals are connected, the GUARD
terminal should be referred to the source of common mode voltage, see Figure 4.
This minimizes errors caused by common mode currents in the measuring circuit
by providing a separate common mode current path.
R1, C1 = Input impedance
R2, C2 = Input-to-Guard
leakage impedance
R3, C3 = Guard-to-Case
leakage impedance
Vcm1, = Common Mode
Vcm2 Voltages
Icm = Common Mode
Current
Icm
Loop
adj061f.emf
Figure 4. External Guard Connections
AC Voltage
The AC Voltage function provides 2-wire measurements using the V INPUT HI
and LO terminals. Push to use the AC Voltage (ACV) function. The
Product makes true-rms ac voltage or ac+dc voltage measurements using a
proprietary sampling method with up to 10 MHz bandwidth. These ranges are
available:
10 mV to 1000 V, where the 10 mV to 100 V ranges provide 121.2 % overrange.
Full scale is 121.2 % of range for these ranges. For example, the 1 V range can
show up to 1.212 V. The 1000 V range can measure up to 1050 V rms.
Input impedance is selectable from 10 MΩ, 1 MΩ, or Auto when dc coupled. Auto
selects the highest impedance available.
24


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8558A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8558A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8558A

Fluke 8558A User Manual - German - 134 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info