Reference Multimeter and 8 ½ Digit Multimeter
Functions
19
When you select Manual, use the softkeys and the numerical keypad to edit the
integration time by PLC and Time. The smallest time aperture is 0 seconds with
200 ns increments and has an upper time limit of 10 seconds.
PLC refers to Power Line Cycles. A PLC at a 50 Hz line is 20 ms; a PLC at a
60 Hz line is 16.67 ms. The smallest aperture that can be set by PLC is 0.01. The
upper limit is the PLC equivalent of 10 seconds so is determined by the line
frequency setting in the Instrument Setup menu. For a 50 Hz line setting, the
maximum is 500 PLC, for a 60 Hz setting it is 600 PLC.
When the aperture is set by time, the display shows the nearest equivalent PLC
to 0.01 PLC precision. When the aperture is set by PLC, the display shows the
aperture in seconds with 200 ns resolution.
Use the navigation keys and to choose the aperture setting method. The
aperture settings for Auto and Auto Fast for different resolution settings are
shown in Table 8.
Measure DC Voltage
The sections below explain how to accurately measure dc voltage.
Simple Lead Connections
For the majority of applications, the simple lead connection without external
guard is adequate as in Figure 2. Use and then (Ext. Guard OFF).
See Input Terminal Selection (INPUTS). The disadvantage of this arrangement is
that the lead connection can form a loop. If a stray alternating magnetic field
(from the line transformer of a neighboring instrument, for example) passes
through the loop, it behaves as a single-turn secondary winding inducing
unwanted ac voltages into the measuring circuit. Use a twisted-pair to reduce the
loop area and adjacent twists will cancel any induced voltages. If you encounter
problems with stray pick-up, Fluke Calibration recommends that you use a
shielded twisted-pair cable with the screen connected to the INPUT LO terminal
at the source as shown in Figure 3.
adj059f.emf
Figure 2. Simple Lead Connections
adj060f.emf
Figure 3. Twisted-Pair Cable Connections
23


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8558A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8558A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8558A

Fluke 8558A User Manual - German - 134 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info