8588A/8558A
Operators Manual
26
Measure DC Current
The Product measures current with the INPUT A and INPUT LO terminals. The
current should flow from the source’s high terminal into the multimeter A terminal
and back to the source’s low terminal out of the multimeter LO terminal.
Similar connection considerations are required for dc current measurement as for
dc voltage measurement. Use shielded twisted-pair cable to reduce induced
interference signals, and connect GUARD to the source of common-mode
voltage to provide a separate common-mode current path.
XW Warning
HIGH CURRENT FLOW
To prevent possible electrical shock, fire, or personal injury,
Do not exceed the Measurement Category (CAT) rating of
the lowest rated individual component of a Product, probe,
or accessory.
Only use probes, test leads, and accessories that have the
same measurement category, voltage, and amperage ratings
as the Product. High current can cause excessive heating of
underrated conductors and may cause a fire.
Note
The current path between the Product terminals is not made when
the current functions are not in use or when front or rear terminals
are deselected.
The rear input terminals may be used to measure currents up to 2 A
only. The rear input A terminal does not share the front panel
automatic protection circuitry, and is instead protected by a fuse
mounted on the rear panel.
Maximum input current capability and protection: The front input
terminals may be used to measure currents up to 30.2 A with
protection for all ranges up to 30.2 A. The front input A terminal
protection for current ranges 1 A and lower has an overload
protection feature if the input significantly exceeds full range. This
protection is automatic and self-resetting, and does not interrupt
current flow. It remains engaged for 1 second after the overload is
removed to minimize circuit interaction and relay reactivation.
WCaution
Damage will occur if >30.2 A is applied to the front current
terminals and the current source maximum compliance is >5 V.
30


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 8558A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 8558A in the language / languages: English as an attachment in your email.

The manual is 1,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 8558A

Fluke 8558A User Manual - German - 134 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info