Documenting Process Calibrator
Source Mode
61
Note
When the Product is set to display temperatures
in Kelvin, the drywell readout will show Celsius,
and when the Product shows Rankine, the
drywell will show Fahrenheit.
Source Scale
This feature scales the output in accordance with the
input requirements of an applicable process instrument’s
response. Percent of scale can be used for linear-
responding transmitters, or square-root responding
transmitters.
Linear-Responding Transmitters
1. If necessary, push for SOURCE mode.
2. Select a source function (, , h, q, t, or
p) as previously described, and record a value.
3. Push the Scale softkey.
4. Select % from the list.
5. Push .
6. Use the numeric keypad to record the 0% of scale
value (0% Value).
7. Push .
8. Use the numeric keypad to record the 100% of scale
value (100% Value).
9. Push the Done softkey.
Percent of scale stays in effect until you change to a
different source function or until the Scale softkey is
pushed and you select a different scale mode.
Square-Root Process Variables
When you select within scaling, the Product output
value is the percent value recorded, squared, and
converted to engineering units.
1. If necessary, push for SOURCE mode.
2. Select a source function (, , h, q, t, or
p) as previously described, and record a value.
3. Push the Scale softkey.
4. Select from the list.
5. Use the numeric keypad to record the 0% of scale
value (0% Value).
6. Push .
7. Use the numeric keypad to record the 100% of scale
value (100% Value).
8. Push .
9. Push the Done softkey.
73


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 754 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 754 in the language / languages: English as an attachment in your email.

The manual is 5,57 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 754

Fluke 754 User Manual - German - 130 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info