Documenting Process Calibrator
Source Mode
59
Source Temperature with a Hart Scientific Drywell
The Product can source temperature using a Hart Scientific
Drywell. Many models are supported.
The drywell driver is able to talk to other drywells from Hart
Scientific, provided that they respond to Hart Scientific’s
standard serial interface commands.
Connect the Product to the drywell by plugging the drywell
interface cable into the pressure module connector as
shown in Figure 24. If the drywell has a DB9 connector, plug
the drywell interface cable directly into the drywell using the
DB9 Null Modem adapter. Drywells with the 3.5 mm jack
connector need to use the serial cable supplied with the
drywell in addition to the Product drywell interface cable.
Join the DB9 connectors of the two cables, and connect the
3.5 mm jack to the drywell.
Be sure the drywell is configured for serial communication at
2400, 4800, or 9600 bits per second. Other rates are not
supported by the Product.
To source a temperature with a drywell:
1. If necessary, push for SOURCE mode.
2. Push t to display the temperature mode menu.
3. Select Drywell from the list of options, and press .
4. The Product will begin to search for a drywell. If the
Product shows Attempting connection for more than
10 seconds, double check your cable connections and
drywell configuration.
5. If a dual well is recognized, a menu will pop up that
allows you to select a "hot" or "cold" side of the dual
well. Only one side of the drywell may be controlled at a
time. Switching sides requires the drywell to be
reconnected, by disconnecting the serial cable or by
leaving drywell source mode and selecting it again.
6. When the drywell is connected, the primary display will
show the actual temperature of the drywell, as
measured by the drywell internally. The drywell model
number will appear above the primary reading. The
setpoint for the drywell is displayed in the secondary
display, at the bottom of the screen. Initially, the
setpoint will be set to the value already stored in the
drywell.
7. Enter the temperature you wish to source and press
.
The settled indicator will be cleared when the actual
temperature is within one degree of the setpoint, and the
actual temperature is not changing quickly. Refer to the
drywell documentation for that model's recommendations for
stabilization time.
The upper temperature limit is restricted by the "High Limit"
setting stored in the drywell. If the Product will not set the
drywell to temperatures within the drywell spec, refer to the
drywell manual to check the "High Limit" setting.
71


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 754 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 754 in the language / languages: English as an attachment in your email.

The manual is 5,57 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 754

Fluke 754 User Manual - German - 130 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info