True-rms Multimeter
Making Basic Measurements
11
Measuring Frequency
XW Warning
To prevent electrical shock, disregard the bar graph for frequencies >1 kHz. If the
frequency of the measured signal is >1 kHz, the bar graph and Z are unspecified.
The Meter measures the frequency of a signal by counting the number of times the signal
crosses a trigger level each second. The trigger level is 0 V, 0 A for all ranges.
Press g to turn on or turn off the frequency measurement function on and off. Frequency
works with ac functions only.
In frequency, the bar graph and range annunciator indicate the ac voltage or current present.
Select progressively lower ranges using manual ranging for a stable reading.
Making Low Impedance Capacitance Measurements
For making capacitance measurements on cables with ghost voltage:
1. Hold q as you turn on the Meter to enable the low-input impedance Capacitance mode.
2. Wait until LCAP shows on the display.
In this mode, capacitance measurements will have a lower accuracy and lower dynamic
range.
Note
This setting is not saved when the Meter is turned off or goes into sleep mode.
V ac Hz
Hz
Hz
15


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Fluke 116 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Fluke 116 in the language / languages: English as an attachment in your email.

The manual is 1,41 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Fluke 116

Fluke 116 User Manual - German - 22 pages

Fluke 116 User Manual - Dutch - 22 pages

Fluke 116 Quick start guide - All languages - 2 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info