XRCISE LINE MED
Technical and visual modifications as well as misprints reserved - © 2019 by ERGO-FIT GmbH & Co. KG
9 5
The displayed diastolic values are too low when active
(With a cuff pressure below the diastole, the microphone picks up pulse wave noise or
noise and interprets it as Korotkoff sound)..
Methodological/basic error:
-tube fails during the measurement process,
-arm movement is too strong,
-other loud noise sources in the area
---> The cause of the noise must be looked into.
Cuff microphone is not positioned accurately. ---> See chapter 7.6.1.
The patient has physiologically very quiet Korotkoff sounds; it is also very difficult
to take their measurements with a stethoscope .
----> Measure on the other arm or re-position the microphone.
Pressure release rate is clearly too high.
-----> Reduce the pressure release rate. (Only possible with BP service program)
⊗ See "No readings at rest".
----> If the control panel displays no values, none can be transmitted
Data transfer emulation is not compatible.
---> Try another combination of protocol selection in the PC ECG and on the
control panel of the machine.
Software error.
----> Try to isolate, understand and document the error. Provide detailed
information about the error to the manufacturer/service partner/software
developer.
Note: If the BL-6 is integrated into the ECG system, use only the ECG software to start
the blood pressure measurement process and not the BP button on the control
panel
When active, systolic values are too low and diastolic values are too high
(The Korotkoff noise is too low and sounds too quiet)
Blood pressure device was remotely started from the PC, but no measured values were
101


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Cardiowise XRCISE Line at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Cardiowise XRCISE Line in the language / languages: English as an attachment in your email.

The manual is 5,31 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Cardiowise XRCISE Line

Cardiowise XRCISE Line User Manual - German - 148 pages

Cardiowise XRCISE Line User Manual - French - 144 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info