BENNING ST 750 A
10.6 RInsu – Insulating Resistance
The insulating resistance is measured between the active parts and each accessible conductive
part including the protective conductor. Moreover, it is measured between the active parts of a
primary circuit and the active parts of the SELV / PELV circuit. It is also measured between the
mains supply unit and all patient connections as well as between patient connections and accessi-
ble conductive parts including the protective earth. In the start menu, select the corresponding VDE
regulation and then the protection class.
Inthe"Testacc.toVDE....(Class.)",selectionmenu,press<RInsu-1>, <RInsu-2> and/or <RInsu-3>.
Start measurement by pressing <Start>.
By means of the setup program, it is possible to set the testing voltage of the device to lower values
(50-500 V DC, beyond the standard), e.g. if test samples are provided with integrated overvoltage ar-
resters. The corresponding nominal and actual values of the testing voltage are shown on the display.
In the current issue of the DIN VDE 0751-1 standard, the measurement of the insulating
resistance is required, if it is considered to be convenient. Insulation measurement must
not be carried out, if it is excluded in the accompanying documents according to the
manufacturer'sspecications.
Limiting valuesfor the insulating resistance are not speciedintheDINVDE0751-1standard
(issue 2008). Here, it is possible to refer to recommendations of the manufacturer or to previously
measured values.
The limiting values according to VDE 0701-0702 or to VDE 0751-1 are taken into consideration.
Finish or stop the measurement by pressing <Stop>, press <Pause>/<Next> for a measuring
pause and press <Reset> to delete the measured value.
Attention! Pressing the <Pause> button will stop the measurement. During the pause,
the test socket will remain live!
Besidesthemeasuringresult (measured value), the overall result is displayed (e.g."Pass"bya
greeneldand"Fail"byaredeld).
Accessible conductive parts which are not connected with the protective conductor must be
scanned by means of the probe / probe tip (VDE 0701-0702; 0751-1).
10.7 IPE – Protective Conductor Current
The protective conductor current measurement, a leakage current measurement carried out by
means of the differential current measurement method "Diff", the direct measurement method
"Dir.",thealternativeleakagecurrentmeasurementmethod"Alt."andthedifferentialcurrentmea-
surementmethodwith current clamp "Clamp" (DIN VDE 0404-4), corresponds to thesumofall
currents running through the insulation of a device to the protective conductor. For the differen-
tialcurrentmeasurementmethodwithcurrentclamp"Clamp",themeasuredvaluesofthecurrent
clamphavetobeenteredunder"Input"!
In the start menu, select the corresponding VDE regulation and then the protection class. In the
"Testacc.toVDE....(Class.)"menu,selectionmenu,pressthe"IPE"button.Selectthedesired
measurementmethod("Diff.","Dir.","Alt."or"Clamp")andstartmeasurementbypressing<Start>.
The limiting values according to VDE 0701-0702 are taken into consideration.
28


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 750A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 750A in the language / languages: English as an attachment in your email.

The manual is 7,87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning ST 750A

Benning ST 750A User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info