10.4 Manual Tests according to VDE 0701-0702 or VDE 0751
Onthemainscreen,theindividualtests/measurements"Testacc.toVDE0701-0702"or"Test
acc.toVDE0751"canbeselectedbypressingthecorrespondingbutton.
Afterhavingenteredtheprotectionclass,the"VisualInspection"menuisdisplayed.Conrmwith
<OK> to continue with the test! For manual measurement, the individual tests / measurements can
be carried out in any order. Normally, however, the order of the individual measurements (tests) has
tobeobservedabsolutely(seechapter11,"StandardTestProcedures"),becausethemeasurement
has to be stopped immediately in case of a failed test. Only after troubleshooting, the test can be
repeated right from the beginning.
Forthersttest/measurement,thetesterstartsaconnectiontesttodetermine,whetherdangerous
contact voltage is applied, whether a test sample is connected and that there is no short-circuit. After
a positive connection test, the measurement is started.
In case of a negative connection test (e.g. open switch), an information message is displayed.
Anotherconnectiontestwillonlybestartedwhenthe"VisualInspection"menuhasbeenpassed
through once again. The test duration is determined by the time between the actuations of the
<Start>/ <Stop> button. An exception is the measurement of the protective conductor resistance
with 10 A. Here, the switch-on time is automatically limited to 5 seconds due to thermal reasons.
By pressing the <Pause> button, it is possible to interrupt measurement so that e.g. the test cable
with the alligator clip or probe tip can be connected to another measuring point without resulting in
incorrect measurements / indications.
Attention! Pressing the <Pause> button will stop the measurement. During the pause,
the test socket will remain live!
Incorrect measured values can be deleted by pressing the <Reset> button. Press <Setup> to get
tothesetupscreen.Onthisscreen,itispossibletomakealldevicesettings(seechapter7,"Com-
missioning/Setup").
Press <About> to call the help function.
10.4.1 Devices with PE (Protection Class I)
Protection class I: Devices with protective conductor connection
Menu, VDE 0701-0702 (protection class I) Menu, VDE 0751 (protection class I)
Selection of individual measurements (tests according to VDE 0701-0702):
RPE Protective conductor resistance ICont Contact current
RInsu-1 Insulating resistance LN–PE Funct. Functional test
RInsu-2 Insulating resistance sec.–PE Cable Continuity test
RInsu-3 Insulating resistance LN–sec. Ua Safety extra-low voltage
IPE Protective conductor current
Selection of individual measurements (tests according to VDE 0751):
RPE Protective conductor resistance PLeak Patient leakage current
RInsu-1 Insulating resistance prim.–PE Funct. Functional test
RInsu-2 Insulating resistance sec.–PE Cable Continuity test
RInsu-3 Insulating resistance prim.–sec.
ILeak Device leakage current
10.4.2 Devices without PE (protection class II)
Protection class II: Devices without protective conductor connection
Menu, VDE 0701-0702 (protection class II) Menu, VDE 0751 (protection class II)
26


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 750A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 750A in the language / languages: English as an attachment in your email.

The manual is 7,87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning ST 750A

Benning ST 750A User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info