08/ 2012 25 / 58
BENNING ST 750 A
- the accessible conductive parts which are not connected to the protective conductor (protection
classes I and II) and the active conductors (mains supply unit)
- the patient connections of applied parts and the active conductors (mains supply unit)
- all patient connections of applied parts of type F and the protective earth of device of protection
class I
- all patient connections of applied parts of type F and accessible conductive parts (which are not
connected to earth) for devices of protection classes I and II.
The test sample must be free of external voltage (disconnected from the mains). For this, contacts of
switches, relays and controlling means in active conductors must be closed (monitoring of the entire
test sample and not only of the supply line). Connect the mains plug to the test socket of the appli-
ancetester.If necessary ,use themeasuring jacks"PE"and "L/N"which areconnectedin parallel
to the"test socket". For test sampleswith accessible conductiveparts (which arenot connected
to the protective conductor), the test must be carried out by means of the additional test cable and
the safety probe tip. For this purpose, connect the test cable to the yellow jack (probe) and scan
all conductive parts of the test sample by means of the safety probe tip. The protective conductor
current, contact current, device leakage current and patient leakage current can be measured by
means of the alternative leakage current measurement method.
1
2
4
mA
PE L1
N
i s ol iert e
Aufst ell ung
1kOhm
3
AP
AP1
AP2
G
~
2
Sonde
PE- Buchse
ST 750A
Ablei t stro m vom Anwe ndungsteil , be i
geräte eigener S tromvers orgung -
direkte Messung:
DIN VDE 07 51-1
ST 750A Gerätetester
1 Stro mmessge rät (mA )
2 Prüfst eckdose, Prüfleitu ng
3 Buchse, „Sonde“
4 P rüflin g
10.3. 3. A b leitstro m messung/Ersa t zableitstro m messung (E rs)
Eine dir ekte Ableit strommessung i st ni cht immer mö glich, wei l dazu die
Geräte entweder isoliert aufgestellt oder an eine von Erde isolierte
Spa nnungsquell e angeschl o sse n w erd e n müss en. Dahe r wird e ine Ersatz-
Able i tstro mmessung durchgef ührt. M essung erfo l gt m it ca. 200V AC. Für
M e s sungen nac h D IN VD E 070 1-0702 gilt eine andere Messanordn ung als
nach DI N VDE 07 5 1- 1 (abgel e itet nac h IEC 60601-1). Die zutref fende
Messanordnung wird entsprechend der Vorauswah l (DIN VDE 0701-0702
oder 07 51) auto matisch e i nges c halt et.
Der ge messene A blei t strom w i r auf die nominel le Netzs pannung (2 3 0V)
hochge rechnet.
Der Ersatzableit strom muss gemessen werden zwischen:
• dem Schutzleit er (berührbare Metalltei le die mit dem Schutzleit er
verbunden sin d, Schutzklasse I) und den aktiven Leiter n (N etz t eil) .
• den berühr baren leitfähigen Teilen die nicht mit dem Schutzleiter
verbunden sin d der Klasse I bis II und den aktiven Leitern (Netztei l).
• den Patienten anschlüssen der Anwendungst eile und den aktiven
Leite rn (Netz teil).
• alle n Pati enten anschlüssen der Anwendungsteile des T yps F und
Schutzerde bei Geräten der Schutzklasse I
• a llen Patient enanschlüssen der Anwendungst eile des Typs F und
(nich t gee r dete n) b erü hr ba r en lei tfähigen T eile n bei Gerä t e n der
Schu t zklas se I und II .
Der Prüf l ing mus s fremdspannungsfrei sein (vom Netz getrennt) . Kontakte
von Sc haltern, R elais und Regele inrichtunge n in aktive n Leitern müssen
dabei geschlossen sei n (Überwachung des gesamt en Prüflings und nicht
nur der Zuleit ung). Zur Aufnahme des Netzstecker s dient die Prüfsteckdose
am Gerätetester . Gegebenenfalls sind die der “Prü fsteckdose”
parallelgeschaltet en Messbuchsen “P E” und “L/N” zu verwenden. B ei
Prüflin gen mit berü hrbaren leit fähigen T eilen (die ni cht mit dem
Schu t zleite r verb unden sind ) muss mit der zusä t zliche n Prüflei t ung und
Sicherheit sprüfspitze geprüft werden. Hierzu wird die Prüfleitu ng in die
gelbe Buchse ( Sonde) eingesteckt und alle leit fähigen Teile des Prüflin gs
werden mit der Sicherheitsprüfspitze abget astet. Schutzleiter strom-,
Berü hrungsstr o m-, Geräteableitstrom und Patientenableit strom-
M e s sungen könne n mit dem E r satz ableitst r ommes sverfahre n durchgef ührt
werden.
1
L N
2
PE
3
mA
G
2kOhm
~
ST 750A
Schu t zleite r stro mmessung / Ersa t z-
Able i tstro m ve rf ahre n:
DIN VDE 07 01- 0702 (Sk I)
ST 75 0A Ger ät etest er
1 Stromm essger ät (mA )
2 Prüfsteckdo se, Prüfleitun g
3 Prüflin g
L1
N
2
4
3
MP
AP
PE
Sonde
1
mA
G
1kOhm
~
ST 750A
Geräte ableitstro m / Ersatz-
Ableit strommessu ng:
DIN VDE 07 51- 1 (SK I)
ST 75 0A Ger ät etest er
1 Stro mmessge rät (mA )
2 Prüfst eckdose, Prüfleitu ng
3 Buchse, „Sonde“
4 Prüfling
Protective conductor current measurement/
alternative leakage current measurement
method:
DIN VDE 0701-0702 (protection class I)
BENNING ST 750 A Appliance T ester
1 current measuring instrument (mA)
2 test socket, test cable
3 test sample
1
2
4
mA
PE L1
N
i s ol iert e
Aufst ell ung
1kOhm
3
AP
AP1
AP2
G
~
2
Sonde
PE- Buchse
ST 750A
Ablei t stro m vom Anwe ndungsteil , be i
geräte eigener S tromvers orgung -
direkte Messung:
DIN VDE 07 51-1
ST 750A Gerätetester
1 Stro mmessge rät (mA )
2 Prüfst eckdose, Prüfleitu ng
3 Buchse, „Sonde“
4 P rüflin g
10.3. 3. A b leitstro m messung/Ersa t zableitstro m messung (E rs)
Eine dir ekte Ableit strommessung i st ni cht immer mö glich, wei l dazu die
Geräte entweder isoliert aufgestellt oder an eine von Erde isolierte
Spa nnungsquell e angeschl o sse n w erd e n müss en. Dahe r wird e ine Ersatz-
Able i tstro mmessung durchgef ührt. M essung erfo l gt m it ca. 200V AC. Für
M e s sungen nac h D IN VD E 070 1-0702 gilt eine andere Messanordn ung als
nach DI N VDE 07 5 1- 1 (abgel e itet nac h IEC 60601-1). Die zutref fende
Messanordnung wird entsprechend der Vorauswah l (DIN VDE 0701-0702
oder 07 51) auto matisch e i nges c halt et.
Der ge messene A blei t strom w i r auf die nominel le Netzs pannung (2 3 0V)
hochge rechnet.
Der Ersatzableit strom muss gemessen werden zwischen:
• dem Schutzleit er (berührbare Metalltei le die mit dem Schutzleit er
verbunden sin d, Schutzklasse I) und den aktiven Leiter n (N etz t eil) .
• den berühr baren leitfähigen Teilen die nicht mit dem Schutzleiter
verbunden sin d der Klasse I bis II und den aktiven Leitern (Netztei l).
• den Patienten anschlüssen der Anwendungst eile und den aktiven
Leite rn (Netz teil).
• alle n Pati enten anschlüssen der Anwendungsteile des T yps F und
Schutzerde bei Geräten der Schutzklasse I
• a llen Patient enanschlüssen der Anwendungst eile des Typs F und
(nich t gee r dete n) b erü hr ba r en lei tfähigen T eile n bei Gerä t e n der
Schu t zklas se I und II .
Der Prüf l ing mus s fremdspannungsfrei sein (vom Netz getrennt) . Kontakte
von Sc haltern, R elais und Regele inrichtunge n in aktive n Leitern müssen
dabei geschlossen sei n (Überwachung des gesamt en Prüflings und nicht
nur der Zuleit ung). Zur Aufnahme des Netzstecker s dient die Prüfsteckdose
am Gerätetester . Gegebenenfalls sind die der “Prü fsteckdose”
parallelgeschaltet en Messbuchsen “P E” und “L/N” zu verwenden. B ei
Prüflin gen mit berü hrbaren leit fähigen T eilen (die ni cht mit dem
Schu t zleite r verb unden sind ) muss mit der zusä t zliche n Prüflei t ung und
Sicherheit sprüfspitze geprüft werden. Hierzu wird die Prüfleitu ng in die
gelbe Buchse ( Sonde) eingesteckt und alle leit fähigen Teile des Prüflin gs
werden mit der Sicherheitsprüfspitze abget astet. Schutzleiter strom-,
Berü hrungsstr o m-, Geräteableitstrom und Patientenableit strom-
M e s sungen könne n mit dem E r satz ableitst r ommes sverfahre n durchgef ührt
werden.
1
L N
2
PE
3
mA
G
2kOhm
~
ST 750A
Schu t zleite r stro mmessung / Ersa t z-
Able i tstro m ve rf ahre n:
DIN VDE 07 01- 0702 (Sk I)
ST 75 0A Ger ät etest er
1 Stromm essger ät (mA )
2 Prüfsteckdo se, Prüfleitun g
3 Prüflin g
L1
N
2
4
3
MP
AP
PE
Sonde
1
mA
G
1kOhm
~
ST 750A
Geräte ableitstro m / Ersatz-
Ableit strommessu ng:
DIN VDE 07 51- 1 (SK I)
ST 75 0A Ger ät etest er
1 Stro mmessge rät (mA )
2 Prüfst eckdose, Prüfleitu ng
3 Buchse, „Sonde“
4 Prüfling
Device leakage current / alternative leakage
current measurement method:
DIN VDE 0751-1 (protection class I)
BENNING ST 750 A Appliance T ester
1 current measuring instrument (mA)
2 test socket, test cable
3 "Probe"jack
4 test sample
L1
N
2
4
3
MP
AP
Sonde
1
mA
G
1kOhm
~
ST 750A
Geräte ableitstro m /
Ersat zableitstro mmessung:
DIN VDE 07 51- 1 ( SK II)
ST 7 5 0A Ge rätete ster
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
L1
N
2
4
3
MP
AP
PE
Sonde
1
mA
G
1kOhm
~
AP1
AP2
ST 750A
Ablei t stro m vom Anwe ndungsteil d es
Typs F – Er satzablei tstr ommessun g:
DIN VDE 07 51- 1 (SK I)
ST 750A Gerä teteste r
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
L1
N
2
4
3
MP
AP
Sonde
1
mA
G
1kOhm
~
AP1 AP2
ST 750A
Ablei t stro m vom Anwendungstei l des
Typs F – Er satzablei tstr ommessun g:
DIN VDE 07 51- 1 ( SK II)
ST 75 0A Ger ät etest er
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
Hinweis: Ersa tzable i tstrom messunge n s ind n u r erl a ubt (V DE 0701-07 02),
wenn die vorheri ge Isolationswiderstandsm essung bestanden wurde! Die
Ersatzmessun g kann zu Ergebnissen führen , die die zulässigen Werte
übers chreiten (v erursac ht z.B. d urch Ablei t konde nsatoren im Prüfli ng ),
dann ist z.B. die Dire kt- oder D ifferenz strommess ung durchz uführen.
10.4. Manuelle Prüfunge n na ch VDE 0701-070 2 o der VDE 0751
Aus de m Hauptbil d sch i rm k ö nnen d ie einzelne n Prüfung e n / M essungen
„Prüfung nac h VDE 070 1-0702“ od e r „Prüfung n a c h VDE 075 1“ per
Tastendruck au fgerufen werden.
→
Nach Eingabe der Sch utzklasse wir d das Menü „Si chtkontro lle“
eingeblen det. Erst nach Eingabe einer Antwort <Ok> ka nn in der P r üfung
fortgef ahren werden. Bei der manuellen Messung k ann die Reihenfolge
der einzelnen Prüfungen / Messungen beliebi g gewählt werden. Jedoch
ist d i e Re i henfo lge der einz elnen M e ssunge n (Prüfunge n) unbedi ngt
einzu halten (s i ehe Ka pitel 11, „ N or m-Prüfabläu fe“), da bei e iner nicht
bes t andene n Prüfung s o for t abge b roc he n werd en muss. E rst na ch
Fehlerbeseit igung darf die Prüfung von Beginn an wiederholt werden.
Das Pr ü fge r ät sta rt et b e i de r erste n Prüfung / M essung e i nen Ans chlusst e st
um zu e rmitteln, o b gefä hrliche Be rührungss pannung a nliegt, e i n Prüfl ing
angeschlo ssen ist un d kein Kurzsch luss vorli egt. Nach posit iver Prüfu ng wird
die Messung ein geschaltet.
Bei ne g ati v em A ns chlusstes t (z. B . Sc ha lter o ffen) wird eine Info einge-
blendet.
Device leakage current / alternative leakage
current measurement method:
DIN VDE 0751-1 (protection class II)
BENNING ST 750 A Appliance T ester
1 current measuring instrument (mA)
2 test socket, test cable
3 "Probe"jack
4 test sample
L1
N
2
4
3
MP
AP
Sonde
1
mA
G
1kOhm
~
ST 750A
Geräte ableitstro m /
Ersat zableitstro mmessung:
DIN VDE 07 51- 1 ( SK II)
ST 7 5 0A Ge rätete ster
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
L1
N
2
4
3
MP
AP
PE
Sonde
1
mA
G
1kOhm
~
AP1
ST 750A
Ablei t stro m vom Anwe ndungsteil d es
Typs F – Er satzablei tstr ommessun g:
DIN VDE 07 51- 1 (SK I)
ST 750A Gerä teteste r
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
L1
N
2
4
3
MP
AP
Sonde
1
mA
G
1kOhm
~
AP1 AP2
ST 750A
Ablei t stro m vom Anwendungstei l des
Typs F – Er satzablei tstr ommessun g:
DIN VDE 07 51- 1 ( SK II)
ST 75 0A Ger ät etest er
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
Hinweis: Ersa tzable i tstrom messunge n s ind n u r erl a ubt (V DE 0701-07 02),
wenn die vorheri ge Isolationswiderstandsm essung bestanden wurde! Die
Ersatzmessun g kann zu Ergebnissen führen , die die zulässigen Werte
übers chreiten (v erursac ht z.B. d urch Ablei t konde nsatoren im Prüfli ng ),
dann ist z.B. die Dire kt- oder D ifferenz strommess ung durchz uführen.
10.4. Manuelle Prüfunge n na ch VDE 0701-070 2 o der VDE 0751
Aus de m Hauptbil d sch i rm k ö nnen d ie einzelne n Prüfung e n / M essungen
„Prüfung nac h VDE 070 1-0702“ od e r „Prüfung n a c h VDE 075 1“ per
Tastendruck au fgerufen werden.
→
Nach Eingabe der Sch utzklasse wir d das Menü „Si chtkontro lle“
eingeblen det. Erst nach Eingabe einer Antwort <Ok> ka nn in der P r üfung
fortgef ahren werden. Bei der manuellen Messung k ann die Reihenfolge
der einzelnen Prüfungen / Messungen beliebi g gewählt werden. Jedoch
ist d i e Re i henfo lge der einz elnen M e ssunge n (Prüfunge n) unbedi ngt
einzu halten (s i ehe Ka pitel 11, „ N or m-Prüfabläu fe“), da bei e iner nicht
bes t andene n Prüfung s o for t abge b roc he n werd en muss. E rst na ch
Fehlerbeseit igung darf die Prüfung von Beginn an wiederholt werden.
Das Pr ü fge r ät sta rt et b e i de r erste n Prüfung / M essung e i nen Ans chlusst e st
um zu e rmitteln, o b gefä hrliche Be rührungss pannung a nliegt, e i n Prüfl ing
angeschlo ssen ist un d kein Kurzsch luss vorli egt. Nach posit iver Prüfu ng wird
die Messung ein geschaltet.
Bei ne g ati v em A ns chlusstes t (z. B . Sc ha lter o ffen) wird eine Info einge-
blendet.
Leakage current of applied parts of type F /
alternative leakage current measurement
method:
DIN VDE 0751-1 (protection class I)
BENNING ST 750 A Appliance T ester
1 current measuring instrument (mA)
2 test socket, test cable
3 "Probe"jack
4 test sample
L1
N
2
4
3
MP
AP
Sonde
1
mA
G
1kOhm
~
ST 750A
Geräte ableitstro m /
Ersat zableitstro mmessung:
DIN VDE 07 51- 1 ( SK II)
ST 7 5 0A Ge rätete ster
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
L1
N
2
4
3
MP
AP
PE
Sonde
1
mA
G
1kOhm
~
AP1
AP2
ST 750A
Ablei t stro m vom Anwe ndungsteil d es
Typs F – Er satzablei tstr ommessun g:
DIN VDE 07 51- 1 (SK I)
ST 750A Gerä teteste r
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
L1
N
2
4
3
MP
AP
Sonde
1
mA
G
1kOhm
~
ST 750A
Ablei t stro m vom Anwendungstei l des
Typs F – Er satzablei tstr ommessun g:
DIN VDE 07 51- 1 ( SK II)
ST 75 0A Ger ät etest er
1 Str omme ssger ät (mA)
2 Prüfsteckd ose, Prüfle i tung
3 Buchse, „ S onde “
4 P rüflin g
Hinweis: Ersa tzable i tstrom messunge n s ind n u r erl a ubt (V DE 0701-07 02),
wenn die vorheri ge Isolationswiderstandsm essung bestanden wurde! Die
Ersatzmessun g kann zu Ergebnissen führen , die die zulässigen Werte
übers chreiten (v erursac ht z.B. d urch Ablei t konde nsatoren im Prüfli ng ),
dann ist z.B. die Dire kt- oder D ifferenz strommess ung durchz uführen.
10.4. Manuelle Prüfunge n na ch VDE 0701-070 2 o der VDE 0751
Aus de m Hauptbil d sch i rm k ö nnen d ie einzelne n Prüfung e n / M essungen
„Prüfung nac h VDE 070 1-0702“ od e r „Prüfung n a c h VDE 075 1“ per
Tastendruck au fgerufen werden.
→
Nach Eingabe der Sch utzklasse wir d das Menü „Si chtkontro lle“
eingeblen det. Erst nach Eingabe einer Antwort <Ok> ka nn in der P r üfung
fortgef ahren werden. Bei der manuellen Messung k ann die Reihenfolge
der einzelnen Prüfungen / Messungen beliebi g gewählt werden. Jedoch
ist d i e Re i henfo lge der einz elnen M e ssunge n (Prüfunge n) unbedi ngt
einzu halten (s i ehe Ka pitel 11, „ N or m-Prüfabläu fe“), da bei e iner nicht
bes t andene n Prüfung s o for t abge b roc he n werd en muss. E rst na ch
Fehlerbeseit igung darf die Prüfung von Beginn an wiederholt werden.
Das Pr ü fge r ät sta rt et b e i de r erste n Prüfung / M essung e i nen Ans chlusst e st
um zu e rmitteln, o b gefä hrliche Be rührungss pannung a nliegt, e i n Prüfl ing
angeschlo ssen ist un d kein Kurzsch luss vorli egt. Nach posit iver Prüfu ng wird
die Messung ein geschaltet.
Bei ne g ati v em A ns chlusstes t (z. B . Sc ha lter o ffen) wird eine Info einge-
blendet.
Leakage current of applied parts of type F /
alternative leakage current measurement
method:
DIN VDE 0751-1 (protection class II)
BENNING ST 750 A Appliance T ester
1 current measuring instrument (mA)
2 test socket, test cable
3 "Probe"jack
4 test sample
Note: According to VDE 0701-0702, alternative leakage current measurements are only permitted,
if the previous insulating resistance measurement has been passed! The alternative measurement
might lead to results which exceed the admissible limiting values (e.g. caused by bypass capacitors
in the test sample). In this case, use the direct or differential current measurement method.
probe
probe
probe
probe