577368
13
Zoom out
Zoom in
Previous page
1/58
Next page
08/ 2012 13 / 58
BENNING ST 750 A
Go to the start menu and press <Setup> / <Expert Settings> to open the menu for entering the
password. Select <Test Procedures>. Then, select the corresponding <VDE Standard…>. Select
the device type (<Devices with PE (Class I)>, <Devices without PE (Class II)>, <SELV (Class
III)>).Allpredenedandcustomizedtestproceduresarelistedinthedrop-downlistbox.Keepthe
last number in mind to save the new test procedure under the next number (15 test procedures
are permanently stored internally). To copy, modify or delete a test procedure, the respective test
procedure must be selected (highlighted in blue)!
1. Creating a new test procedure:
After pressing <New Test Procedure>, enter a new number and name in the upper
field. If necessary, a pre-setting can be corrected by means of the <<--> button or
deleted by means of the <Clear> button. Select / enter the desired test parameters
in the main field. Press <Limit Values> to modify the limiting values (see chapter
7.1"Limiting Values") . After pressing <Back> in the "Test Procedure Setup" menu,
press <Save> / <Yes> to register a new test procedure in the drop-down list box.
Press <Back> to return to the start menu.
2. Copying a test procedure:
Select the test procedure to be copied and then press <Copy Test Procedure>. The further
proceeding is analogous to creating a new test procedure.
3. Modifying a test procedure:
Select the test procedure to be modified and then press <Modify Test Procedure>. The further
proceeding is analogous to creating a new test procedure. During the saving process of the
<Modify Test Procedure> option, it is indicated that the data record already exists and you will
be asked, if you want to overwrite it (<Yes> / <No>).
4. Deleting a test procedure:
Select the test procedure to be deleted and then press <Delete Test Procedure>. For the
<Delete Test Procedure> option, you will be asked, if you really want to delete the test
procedure (<Yes> / <No>).
7.5.3 Test Company / Person
Go to the <Expert Settings> menu to create or to modify the <Test Company / Person>. Test
company and test person are stored in the tester and will be saved in the database for each test!
In case of several test persons, implicitly check the settings!
7.5.4 Updating the Firmware
Thermwareisinstalled(update)viatheSDmemorycard.Thecorrespondinglesareprovidedby
BENNING.CopythisleontoanSDmemorycard(max.2GB).Afterhavinginsertedthememory
card,installthermwareupdateunder<Setup> / <Expert Settings> / <Firmware Update> and
conrmtheupdatewith<Yes>. After successful update, switch the device OFF and ON again.
7.5.5 Updating the User Interface
It is possible to install a software update via the USB interface or the SD card slot. The correspon-
dingles(WindowsCE)areprovidedbyBENNINGina directory called "PicoMODX".Copythis
directory onto a USB stick. Switch the BENNING ST 750 A Appliance Tester ON. After inserting the
USB stick into the A jack, switch the BENNING ST 750 A Appliance Tester OFF and ON again. Now,
the device automatically installs the software. Remove the USB stick after completion. After suc-
cessful installation, the system settings in the setup program have to be made again.
For this purpose, successively select the following: <Setup> / <System Settings> / <System
Data>. Now, select e.g. the functions <Barcode Scanner>, <Mains Polarity Reversal ON>,
<Buzzer ON> and <Short-circuit Test ON>. Save the settings with <Save> and return to the start
menu by pressing <Back>.
7.5.6 System Reset
By pressing <Setup> / <Expert Settings> / <System Reset>, all system data will be reset. Con-
rmtheresetbypressing<Yes> / <No>. All limiting values and all user data will be reset to the
factory delivery status.
7.6 Date / Time
Gotothe"Setup Selection"menu,pressthe<Date / Time> button and enter the current values.
Use the input stylus to mark the corresponding numerical values and select the values / numerals.
Afterhavingnishedthesetting,pressthe<Save> button.
13


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 750A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 750A in the language / languages: English as an attachment in your email.

The manual is 7,87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning ST 750A

Benning ST 750A User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info