577368
11
Zoom out
Zoom in
Previous page
1/58
Next page
08/ 2012 11 / 58
BENNING ST 750 A
<Printer>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
7.3.1 System Data
The device version and serial number are listed under <Setup> / <System Settings> /
<System Data>. The duration of the individual test procedures (standard = 5 sec.) is determined by
the test time. The safety shut-off terminates a test procedure (standard = 30 sec.).
Thecontactsurfacesbetweenconnectorshaveonlylowtransitionresistances.Todenethetransi-
tionresistance,herethe assumable resistance value is entered (standard = 1Ω). Moreover,itis
denedwhetheraprinterorabarcodescannerisconnectedtotheserialinterface.
Mains pole reversal ON: On the test socket, L and N are interchanged.
Buzzer ON: Acoustic signal in case of faults
Short-circuit test ON: Whenstartingthetest,rstcheckforshort-circuit.
Standard test procedures:
Automatic test procedures complying with DIN VDE
0701/0702/0752 are activated.
Custom test procedures:
Customized test procedures are activated. Customized test pro-
cedures are stored on the SD memory card.
Test repetition ON:
This mode allows measurements of several measuring points by
means of the test cable (probe). It can be used for measuring
RPW, RISO (protection classes II/III), Icont, Ileak and IPLeak.
The corresponding maximum and minimum value of all measure-
ments will be stored.
Database backup ON:
The device automatically creates up to 5 backups for each data-
base, e.g. Database name_backup.001.
7.3.2 Factory Settings
Select <Setup> / <System Settings> / <Factory Settings> to reset all limiting values to the factory
settings.Conrmyourentrywith<Yes> / <No>!
08/ 2010 10 / 58
BENNING ST 750
RPE data (resistance measurement of the PE conductor):
RPE VDE 701-702 = resistance for a cable length of 5 m and 1.5mm²
RPE VDE 751 = resistance for a cable length of 5 m and 1.5mm²
RPE per 7.5 m line length = additional resistance per 7.5 m line length
Line length = total length of the device supply line
Cross-section = cross-sectional area of the device supply line
RInsu data (insulation measurement):
RInsu-1 VDE 701-702 = insulating resistance of primary to PE
RInsu-1 VDE 701-702 heating elements = insulating resistance of primary to PE
RInsu-2 / RInsu-3 VDE 701-702 = insulating resistance of secondary to PE / primary to secondary
RInsu-1 VDE 701-702 Ua (SELV/PELV) = insulating resistance of primary to PE (SELV devices)
RInsu-1 VDE 751 = insulating resistance primary to PE
RInsu-2 VDE 751 = insulating resistance of secondary to PE
RInsu-3 VDE 751 = insulating resistance of primary to applied part
Testing voltage for primary to PE = testing voltage for primary to PE
Testing voltage for secondary to PE = testing voltage for secondary to PE
Testing voltage for primary to secondary = testing voltage for primary to secondary
IAL data (alternative leakage current measurement):
IPE VDE 701-702 protection class I = PE alternative leakage current
ILeak VDE 701-702 protection class II = alternative leakage current of accessible parts
ILeak VDE 751 protection class I = PE alternative leakage current
ILeak VDE 751 protection class I = alternative leakage current of accessible parts
Patient leakage current BF = alternative patient leakage current of applied part, type BF
Patient leakage current CF = alternative patient leakage current of applied part, type CF
ILeak data (leakage current measurement):
IPE VDE 701-702 protection class I = PE leakage current
ILeak VDE 701-702 protection class II = leakage current of accessible parts
ILeak VDE 751 protection class I = PE leakage current
ILeak VDE 751 protection class I = leakage current of accessible parts
Patient leakage current BF = patient leakage current of applied part, type BF
Patient leakage current CF = patient leakage current of applied part, type CF
ICont data (contact current measurement):
ICont VDE 701-702 = contact current to PE
Functional data (functional measurement):
IPE VDE 701-702 = PE leakage current (differential current measurement)
ICont VDE 701-702 = contact current to PE (differential current measurement)
IPE VDE 751 = PE leakage current (differential current measurement)
ICont VDE 751 = contact current to PE (differential current measurement)
Ua data (measurement on PELV / SELV devices):
Max. output voltage = max. output voltage for PELV / SELV devices
Cable data (cable measurement):
Line length = simple line length
Cross-section = cross-sectional area of the conductor
Conductor qty. = conductor quantity (for total length)
7.2 System Settings
Select <Setup> / <System Settings> to modify and permanently save the baud rate, parity, data
bits and stop bits e.g. for barcode scanners, RFID readers, PCs and printers.
see chapter 16 to 18
<Barcode Scanner>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
<RFID Reader>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
<PC>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
<Printer>
Thefollowingeldsarereleasedfortheuser:
- setting of baud rate, parity, data bits and stop bits
7.2.1 System Data
The device version and serial number are listed under <Setup> / <System Settings> /
<System Data>. The duration of the individual test procedures (standard = 5 sec.) is determined by
the test time. The safety shut-off terminates a test procedure (standard = 30 sec.).
Thecontactsurfacesbetweenconnectorshaveonlylowtransitionresistances.Todenethetransi-
tionresistance,herethe assumable resistance value is entered (standard = 1Ω). Moreover,itis
denedwhetheraprinterorabarcodescannerisconnectedtotheserialinterface.
Mains pole reversal ON: On the test socket, L and N are interchanged.
7.3.3 Language Setting
In the <System Settings> menu, press the <Language Settings> button to select the desired
language. Then, press <Save>.
7.3.4 Entering a Password
AlostpasswordcanonlyberesetbytheBenningServiceteam(seechapter24,"Pro-
ductSupport")!
The expert setting function can only be accessed by means of a password. Upon delivery, the de-
viceisprovidedwiththepassword"Benning".ThepersonresponsiblefortheBENNINGST750A
ApplianceTestershalldeneanewcondentialpassword.
To dene a new password, subsequently press <Setup> / <System Settings> / <Password
Entry> / <Input>. First, the current password is requested. Then, the new password will be as-
signed, compared and saved with <OK>.
11


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 750A at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 750A in the language / languages: English as an attachment in your email.

The manual is 7,87 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Benning ST 750A

Benning ST 750A User Manual - German - 68 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info