BENNING ST 725
31
permanent measurements at a high load current. If the admissible internal operating
temperature is exceeded, the "StOP" and "hot" symbols will be shown on the display. In this
case, the BENNING ST 725 must be disconnected from the mains and can be used again
after a sufficient cooling phase.
Note on measuring the contact current:
- Accessible conductive parts which are not connected with the protective conductor have to
be tested as described in section 9.2. For measuring the contact current (direct measurement
method), the BENNING ST 725 has to be operated with a mains voltage of 230 V.
- During contact current measurement by means of the direct measurement method, no part
of the test object must have a connection to the earth potential. The test object must be
placed onto an insulated surface. Otherwise, leakage currents to earth might influence the
measuring result.
9.2 Testing of devices of protection class II
(shock-proof) and of devices of protection
class III
(safety extra-low voltage)
Testing of devices without protective conductor and with accessible conductive parts
- Connect the test object to the test socket
1
of the BENNING ST 725.
- Establish a connection between the 4 mm test socket
9
and a metal part of the test object
by means of the test lead with alligator clip.
- For mains operating mode (contact current by means of the direct measurement method,
test object in operation!): Connect the plug of the mains connection cable with the socket K
and the shock-proof plug with a protected shock-proof socket (230 V, 50 Hz, 16 A).
- If necessary, the testing voltage of the R
ISO
(insulating resistance) measurement can be
reduced to 250 V
DC
by means of the -key
5
. The selected testing voltage is briefly shown
on the display
8
. Press the key again to switch back to the preset testing voltage of 500 V
DC
.
- Switch the test object on.
- Press the
-key
3
to start the automatic testing procedure.
- If "Lo LOAD" is shown on the display, please check whether the test object is switched on.
- Press the key
3
to continue the testing procedure in case of the load being too low (R
L-N
>
6kΩ).
- If "HIGH LOAD“ is shown on the display, this indicates an excessive load (R
L-N
<<14Ω,I
LAST
(I
LOAD
) > 16 A) of the test object. There might be danger of a m short-circuit or of an earth
fault. Check whether there is a short-circuit between the external conductor (L) and neutral
conductor (N) of the test object.
- If there is no short-circuit, you can continue with the testing procedure by pressing the
-key
3
.
- If the insulating resistance R
ISO
is higher than the admissible limiting value, a appears
next to the R
ISO
symbol.
BENNING ST 725 in mains operating mode:
- The BENNING ST 725 interrupts the testing procedure after the R
ISO
(insulating resistance)
measurement and requests the user to switch the mains voltage of 230 V to the test socket
1
by showing a flashing "I
LEAK
" symbol. Make sure that the test sample is protected and
press the -key
4
to measure the contact current I
LEAK
(direct measurement method).
- The contact current by means of the direct measurement method only starts as soon as the
mains voltage is correct applied.
40


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 725 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 725 in the language / languages: English, German, French as an attachment in your email.

The manual is 10,68 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info