BENNING ST 725
30
measurement and requests the user to switch the mains voltage of 230 V to the test socket
1
by showing a flashing "I
LEAK
" symbol. Make sure that the test sample is protected and
press the -key
4
to measure the protective conductor current by means of the differential
current measurement method.
- The protective conductor current measurement (differential current measurement method)
only starts as soon as the mains voltage is correct applied.
step 1 of 2:
- After a measuring time of 5 seconds or by pressing the key
4
, mains polarity will be
reversed and the protective conductor current will be measured with reversed mains voltage
("L/N" - "N/L"). The highest measured value of both measurements will be displayed. (step 2
of 2)
step 2 of 2:
- If the protective conductor current is lower than the admissible limiting value, a will be
shown next to the I
LEAK
symbol.
- The overall test is considered to be passed, if "PASS" is shown on the display.
As an alternative:
BENNING ST 725 in battery operating mode (without mains supply):
- Similarly, a will be shown next to the I
EA
symbol, if the protective conductor current I
EA
(al-
ternative leakage current measurement method) is lower than the admissible limiting value.
- The test is considered to be passed, if "PASS" is shown on the display.
Seegure4: TestingofdevicesofprotectionclassI(deviceswithprotectiveconductorand
accessible conductive parts which are connected to the protective conductor)
Note on measuring the protective conductor resistance:
- Alternatively, the measurement of the protective conductor resistance R
PE
can be carried out
as permanent measurement (max. 2 x 90 seconds). For this purpose press the key
2
for
approx. > 5 seconds until the symbol appears on the display. Check the connecting line of
the test object by bending it over the entire length in order to detect weak points or a break of
the protective conductor. The BENNING ST 725 continuously records the current measured
value on the display and stores the maximum value in the memory. By pressing the key
2
again, the measurement is carried out with reversed polarity. Press the key
2
again to
indicate the maximum value of R
PE
on the display and to continue the testing procedure as
described in section 9.1.
Note on measuring the protective conductor current in mains operating mode:
- Alternatively, the measurement of the protective conductor current I
LEAK
can be carried
out as permanent measurement (max. 2 x 5 minutes). Press the key
4
for approx. > 5
seconds to start permanent measurement. After 5 minutes, the polarity of the mains voltage
will be reversed automatically ("L/N" - "N/L"). By pressing the key
4
earlier, the mains
voltage polarity reversal can be activated manually and by pressing the key
4
again, the
measurement can be stopped.
Please observe that the BENNING ST 725 is not designed for carrying out repeated
39


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Benning ST 725 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Benning ST 725 in the language / languages: English, German, French as an attachment in your email.

The manual is 10,68 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info