6-3
IM 1F5B2-01E
6. PREPARING FOR OPERATION
6.2 Pressure Loss and Cavitation
When changing the range, ALWAYS make sure the line’s back pressure level will not
lead to adverse effects due to pressure loss or cause cavitation.
Pressure Loss
The pressure loss is calculated by
P=108 × 10
-5
× ρf • υ
2
........................ (1)
or
P=135 × ρf •
Qf
2
D
4
..................................... (2)
where
P = pressure loss (kPa)
Qf = volumetric flow rate under operating conditions (m
3
/h)
D = inner diameter of ULTRA YEWFLO (mm)
υ = flow velocity (m/s)
ρf = density under operating conditions (kg/m
3
)
Cavitation (Minimum Back Pressure of the Line)
IMPORTANT
Cavitation may occur in fluid measurements if the line pressure is too low and the flow
velocity is too high, resulting in failure to measure the correct flow rate. Before
measurement, ALWAYS make sure the line’s back pressure level will not cause
cavitation.
No cavitation occurs if the minimum back pressure of the line at the maximum flowrate
is sufficiently low compared with the actual line pressure.
The minimum line pressure is calculated by
P=3.8 × ∆P + 1.3 × Po ......................... (3)
where
P = line pressure at a point two to seven times the nominal
diameter downstream from the ULTRA YEWFLO (kPa
abs)
P = pressure loss (kPa)
Po = liquid’s saturation vapor pressure under operating
conditions (kPa abs)
6.3 Zero Adjustment
No zero adjustment is necessary since the zero point does not shift.
6.4 Resetting totalized Value
•On the Brain Terminal (BT200), choose EXECUTE from the menu in parameter E10
(TOTAL RESET).
•Or, press the RESET switch located on the indicator/totalizer component.
40


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Yokogawa UYF200 Ultrasonic Vortex Flowmeter at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Yokogawa UYF200 Ultrasonic Vortex Flowmeter in the language / languages: English as an attachment in your email.

The manual is 0,93 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info