<14. EXPLOSION PROTECTED TYPE INSTRUMENT>
14-3
IM 01F06A00-01EN
(3) Operation
WARNING
Wait 3 min. after power is turned off, before
opening the covers.
Take care not to generate mechanical
spark when access to the instrument and
peripheral devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modi cation or parts
replacement by other than authorized
representative of Yokogawa Electric
Corporation is prohibited and will void the
certi cation.
(5) Installation Diagram of Intrinsically safe
(and Note)
DY (Flowmeter)
SUPPLY
PULSE
Safety barriers
[Integral type]
Electrical data:
Signal/Supply Circuit (Terminals SUPPLY + and –):
Ui = 30 V, Ii = 300 mA, Pi = 0.9 W (linear source),
Ci = 14 nF, Li = 0 mH
Pulse Circuit (Terminals PULSE + and –):
Ui = 30 V, Ii = 300 mA*, Pi = 0.9 W (linear source),
Ci = 14 nF, Li = 0 mH
*: Refer to “Contact rating” for the maximum current value of
Pulse Circuit
Note:
- It shall be assured that the voltage difference between the
output circuits of the associated apparatus (safety barriers)
is not more than 30V.
- Cables for the connection for Signal/Supply Circuit and
Pulse Circuit shall be of Type A or B in accordance with EN
60079-14, otherwise the sum of Io of the associated
apparatus (safety barriers) shall be not more than 300 mA.
+
-
+
-
+
-
+
-
+
-
+
-
+
-
+
-
+
-
+
+
-
+
[Remote type without built-in Temperature sensor]
PULSE
A
B
T
C
A
B
T
Safety barriers
DY-N (Detector)
DYC: Signal cable
Non Hazardous
Location
Non Hazardous
Location
F1402.ai
Hazardous
Location
Hazardous
Location
DYA (Converter)
SUPPLY
+
-
+
-
+
-
+
-
+
-
+
[Remote type with built-in Temperature sensor]
PULSE
A
B
T
C
A
B
T
Safety barriers
DY-N (Detector)
DYC: Signal cable
Non Hazardous
Location
Hazardous
Location
DYA (Converter)
SUPPLY
(6) Electrical Connection
The type of electrical connection is stamped near
the electrical connection port according to the
following codes.
Screw size
ISO M20 X 1.5 female
ANSI 1/2-14NPT female
!
!
M
N
Marking
F1403.ai
154


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Yokogawa DYA Vortex Flow Converter at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Yokogawa DYA Vortex Flow Converter in the language / languages: English as an attachment in your email.

The manual is 5,84 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info