Muscle tremor
Causes
The patient is uncomfortable, tense, nervous.
The patient is cold and shivering.
The exam bed is too narrow or short to comfortably support arms and legs.
The arm or leg electrode straps are too tight.
Actions
Verify that the patient is comfortable, warm, and relaxed.
Check all electrode contacts.
If interference persists, turn the muscle-tremor filter on. If interference still persists,
the problem is probably electrical in nature. See the suggestions for reducing AC
interference (in a related troubleshooting tip).
AC interference
AC interference superimposes even-peaked, regular voltage on the waveforms.
Causes
The patient or technician was touching an electrode during recording.
The patient was touching a metal part of an exam table or bed.
A lead wire, patient cable, or power cord are broken.
Electrical devices in the immediate area, or lighting, or wiring concealed in walls or
floors are interfering.
An electrical outlet is improperly grounded.
The AC filter is turned off or set incorrectly.
Actions
Verify that the patient is not touching any metal.
Verify that the AC power cable is not touching the patient cable.
Verify that the proper AC filter is selected.
If interference persists, unplug the electrocardiograph from AC power and run it on
the battery. If this solves the problem, you’ll know that the noise was introduced
through the power line.
If interference still persists, the noise may be caused by other equipment in the
room or by poorly grounded power lines. Try moving to another room.
36 Troubleshooting CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph
40


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Welch Allyn CP 50 Plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Welch Allyn CP 50 Plus in the language / languages: English as an attachment in your email.

The manual is 1,34 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Welch Allyn CP 50 Plus

Welch Allyn CP 50 Plus User Manual - Dutch - 52 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info