Troubleshooting
Lead-quality problems
“Artifact” message on the screen
Artifact is signal distortion that makes it difficult to accurately discern the waveform
morphology.
Causes
The patient was moving.
The patient was shivering.
There is electrical interference.
Actions
See actions for wandering baseline, muscle tremor, and AC interference.
Wandering baseline
Wandering baseline is an upward and downward fluctuation of the waveforms.
Causes
Electrodes are dirty, corroded, loose, or positioned on bony areas.
The electrode gel is insufficient or dried.
The patient has oily skin or used body lotions.
Rising and falling of chest during rapid or apprehensive breathing.
Actions
Clean the patient’s skin with alcohol or acetone.
Reposition or replace the electrodes.
Verify that the patient is comfortable, warm, and relaxed.
If wandering baseline persists, turn the baseline filter on.
35
39


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Welch Allyn CP 50 Plus at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Welch Allyn CP 50 Plus in the language / languages: English as an attachment in your email.

The manual is 1,34 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Welch Allyn CP 50 Plus

Welch Allyn CP 50 Plus User Manual - Dutch - 52 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info