ZIE PAG 21 : SWITCH INDELING
4.1 ADJUSTING : INPUT SENSITIVITY SWITCH 1 - 3
Amplifier factor: 1x . . 8x : De digital amplifier of the 348 may be
adjusted in 8 running up steps of
Table 1 0.25 mV/V each. Setting by switches 1 up to 3, see table 1:
De highest amplification (8 x) is achieved at an input sensitivity of 0.25 mV/V.
Do not set the amplifier at a higher rate than necessary.
The setting is not very critical. Only at more than 15,000 scale divisions
external resolution some noise may occur. The number of divisions of the
selected input range is 15,000 counts, but may run up to ca. 18,000 counts.
At an improper setting of the input range it is possible to exceed the linear
range and/or the calculation limits of the A/D converter.
Exceeding the calculation limits will cause a stop of the A/D converter.
Excess of the input range over 15,000 divisions is shown in the internal display
with a “ ! ” in the first decade. In the data string a > of < appears as an
overload error message in character 4 of the string.
Attention ! The input sensitivity has to be adjusted before all other settings and adjustments !
Calculation example:
Required weighing capacity : 100 kg
Applied loadcells : 3 x 100 kg = 300 kg, 3 loadcells of 2.00 mV/V
Loadcell signal at 100 kg : 100 kg : 300 kg x 2 mV/V = 0.67 mV/V, round off upwards to 0.75 mV/V
Number of divisions : 0.67 mV/V : 0.75 mV/V x 15,000 counts = 13,400 scale divisions available
Scale division will be (1,2 of 5) : 100 kg : 13,400 divisions = 7.5 gram, round off upwards to 10 gram
4.2 ADJUSTING : ZERO SETTINGS SWITCH 4 - 5
Automatic zero setting : Switch 4 “on” sets the so called “zero track” active. This zero track keeps,
with the 348 in operation automatically and continuously in gross and net mode, the zero reading exact
at zero, as soon as the weight measuring deviates >
0.1 scale division from
the zero, with a maximum of 0.5 division and a maximum speed of 0.5 division
per second. Switch 5 too has to be “on”.
The zero track is based upon a separate digital memory, and has no
connection with the dead load compensation of the dil switches 6 - 16.
The range of the zero track is limited within - 1.3% . . . + 2.7% of the weighing
capacity, as set by the overload switches, see chapter 4.5.
These limits are valid too for the external “set zero” command: ! Z .
Automatic zero setting : With switch 5 “on” at switching on the 348, the reading will be automatically set
at switching on the 348 zero, after no motion control (high level S2), with 0.01 division accuracy.
This zero setting is based upon a separate digital memory, and has no
connection with the dead load compensation with the dil switches 6 - 16.
The range of this initial zero setting is at legally stamped scales limited from
- 5% . . + 15% of the weighing range as set by the overload switches.
For not stamped scales (switch 60 “off”) the zero setting range is extended
from -20% . . +80%.
Manual 348-2 V 3.1 Page 13 01-06-2007
Switch Input range
1 2 3 mV/V
0 0 0 0 - 0.25
0 0 1 0 - 0.50
0 1 0 0 - 0.75
0 1 1 0 - 1.00
1 0 0 0 - 1.25
1 0 1 0 - 1.50
1 1 0 0 - 1.75
1 1 1 0 - 2.00
13


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Weightec 348 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Weightec 348 in the language / languages: English as an attachment in your email.

The manual is 1,66 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Weightec 348

Weightec 348 User Manual - Dutch - 49 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info