652497
40
Zoom out
Zoom in
Previous page
1/44
Next page
38
Troubleshooting (continued):
Problem Solution
Stimulation is weak, even at
maximum intensity
This is most likely due to low batteries.
Replace Power Packs. This can also be
caused by a decrease in sensation over
certain areas of the skin (e.g., those without
good nerve conduction, immediately after
ice application, after pain medications are
taken or injected). Use caution when
increasing intensity in these situations.
Screen Displays:
A problem has been
detected with your
device.
The device did not pass the internal
diagnostic testing. Do not use the device.
Call Patient Care for a replacement.
Device beeps and the screen
says:
Incomplete setup.
Please set your
intensity.
The device is still in the set up mode where
you set the intensity for each treatment and
the actual treatment program has not yet
started (there will not be a timer at the
bottom of the screen even though you will
feel the stimulation). Press (P) to Proceed,
then set the intensity for each treatment
segment. The screen will say start counting
down from 5 when you have completed the
set up.
Screen Displays:
Warning. Incorrect
power supply connect-
ed. Please use the AC
Adapter provided by
VQ OrthoCare for
this device.
The SurgiStim4 device is designed to only
work with the VQ supplied AC Adapter. This
screen appears when a different AC Adapter
is used and allows too much current, which
could damage the device. Use the AC
Adapter provided by VQ OrthoCare
for this device.
Screen Displays:
Battery Low. Replace
Power Packs or plug
in the AC Adapter.
The Power Packs are low. Replace with
a new set or plug in the AC Adapter.
40


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for VQ OrthoCare SurgiStim 4 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of VQ OrthoCare SurgiStim 4 in the language / languages: English as an attachment in your email.

The manual is 0,62 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info