622303
6
Zoom out
Zoom in
Previous page
1/53
Next page
DVM9912 VELLEMAN
6
1. Set the function switch to the V position (“mV” will appear on the display). The device will automatically
select DC.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana plug into
the positive (V) jack.
3. Touch the test probe tips to the circuit under test. Be sure to observe the correct polarity (red lead to
positive, black lead to negative).
4. Read the voltage on the display. The display will indicate the proper decimal point and value. If the polarity
is reversed, the display will show (-) minus before the value.
AC VOLTAGE MEASUREMENTS
1. Set the function switch to the V position. Press MODE to select AC voltage.
2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana plug into
the positive (V) jack.
3. Touch the test probe tips to the circuit under test.
4. Read the voltage in the display. The display will indicate the proper decimal point, value and symbol (AC,
V, etc.).
DC CURRENT MEASUREMENTS
1. Insert the black test lead banana plug into the negative (COM) jack.
2. For current measurements up to 4000µA DC, set the function switch to the µA position and insert the red
test lead banana plug into the (µ A) jack.
3. For current measurements up to 400mA DC, set the function switch to the mA range and insert the red test
lead banana plug into the (mA) jack.
4. For current measurements up to 10A DC, set the function switch to the A position and insert the red test
lead banana plug into the 10A jack.
5. Press the MODE button until “DC” appears in the display.
6. Remove the power from the circuit under test, then open up the circuit at the point where you wish to
measure current.
7. Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive
side of the circuit.
8. Apply power to the circuit.
9. Read the current on the display. The display will indicate the proper decimal point, value and symbol.
WARNING:
Risk
of
e
lectrocution. The probe tips may not be long enough to contact the live parts inside
some 240V appliance outlets because the contacts are recessed deep in the outlets. As a
result, the reading may show 0V when the outlet actually carries voltage. Make sure the
probe tips are touching the metal contacts inside the outlet before assuming that there is no
voltage is present.
CAUTION:
Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large
voltage surges may occur that can damage the meter.
CAUTION:
Do not carry out current measurements on the 10A scale for a period longer than 30
seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
6


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Velleman DVM9912 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Velleman DVM9912 in the language / languages: English, German, Dutch, French, Spanish as an attachment in your email.

The manual is 1,16 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info