Calibration
The thermometer is initially calibrated at the time of manufacture.
If the thermometer is used according to the use instruction, periodic
readjustment is not required. However, We recommends checking
calibration every two years or whenever clinical accuracy of the
thermometer is in question. Please send the complete device to the
dealers or manufacturer.
The above recommendations do not supersede the legal requirements.
The user must always comply with legal requirements for the control
of the measurement, functionality, and accuracy of the device which
are required by the scope of relevant laws, directives or ordinances
where the device is used.
A clinical summary and procedures for checking calibration are
available upon request.(Turn on the thermometer and press the power
button long time until entering into calibrate mode, software version
will be displayed.)
Direct Current
Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products
Type B F applied part
Consult Accompanying Documents
LOT Batch Code
26
Troubleshooting
In Ear mode:
Temperature taken is
higher than 43.0
(109.4).
In Object mode:
Temperature taken is
higher than 100
(212).
Problem Solution
Read Temperature
Taking Hints thoroughly ,
then make sure the probe
cover and lens filter are
clean, then take a new
temperature measurement.
The thermometer could
not work due to low
battery.
Replace two new
alkaline batteries
size AAA.
Error message
Read Temperature Taking
Hints Thoroughly, then
check the integrity of the
probe cover and take a new
temperature measurement.
24
Remove probe cover in
object temperature mode.
25
In object temperature
mode, measurement
with probe cover.
In Ear mode:
Temperature taken is
lower than 32.0
(89.6).
In Object mode:
Temperature taken is
lower than 0
(32).
23
Storage and Transportation Temperature Limit:
-13 131(-25~55℃)
-25
55
Specifications
Ear/Object mode:
10~40(50~104), 15%~85%RH, non-condensing
00hPa ~ 1060hPaAtmospheric Pressure : 7
57*146*38mm
Approx.83grams(with batteries)
DC3V(2×AA battery)
Measuring range
Laboratory accuracy
Display resolution 0.1 or 0.1
Measure time Approximately one second
Operating temperature
range:
Storage and transport
temperature range
-25~ 55 (-13~131), 15%~95%RH, non-condensing
00hPa ~ 1060hPaAtmospheric Pressure : 7
Clinical accuracy
Dimension
Weight
Battery
Battery life Approx. 1 year/6000 readings
Measuring site Ear canal(Ear Mode)
Oral (This thermometer converts the ear temperature to
display its “oral equivalent.)
Reference body site
Operation mode Ear mode(Adjust mode)
Ingress protecting rating
IP22
Clinical bias0.08℃(0.14℉)
Clinical repeatability0.13℃(0.23℉)
Limits of agreement:0.73℃(1.31℉)
Shock withstands drop of 3 feet
Expected service life Three years
Ear mode: 32.0~43.0(89.6~109.4)
Object mode: 0~100(32~212)
Ear mode:
±0.2 (0.4) during 35.5~42.0 (95.9~107.6)
at 15~35 (59.0~95.0) operating temperature range
±0.3 (0.5) for other measuring and operating
temperature range
Object mode: ±4% or ±2(4) whichever is greater
Troubleshooting
Problem Solution
The ambient temperature
is not within the range
between 10 and 40
(50~104).
Place the
in a room for at least 30
minutes at room
temperature between
10 and 40
(50~104)
thermometer
The thermometer is
placed incorrectly
or unsteady.
The thermometer
showing a rapid ambient
temperature change.
The thermometer is not
functioning properly.
Read Illustration For
Use thoroughly and
take a new temperature
measurement.
Allow the thermometer
to rest in a room for at
least 30 minutes at
room temperature:
between 10 and 40
(50~104)
Unload the battery,
wait for 1 minute and
repower it. If the
message reappears,
contact the retailer
for service.
Error message
Take a measurement
until or appears
on the display.
Measurement before
thermometer is ready
Product Description
1. Probe Cover
2. Probe
3. Test Button
4. Button
5. Battery Cover
6. Setting Button
3
1
2
5
3
4
6
Electromagnetic Compatibility Information Electromagnetic Compatibility Information
7


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Unknown DET-1015b Infrared Ear Thermometer at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Unknown DET-1015b Infrared Ear Thermometer in the language / languages: English as an attachment in your email.

The manual is 4.36 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info