layer thickness measuring device BB30
Device calibration
Carry out a zero point calibration before each measuring
operation.
You can perform the calibration e.g. at an untreated or uncoated
spot of the object to be tested or use the set included in the
scope of delivery.
Info
Please bear in mind that the device at hand is a
precision measuring device that can determine coating
thicknesses of no more than a few micrometres (1µm
is equivalent to one thousandth of a millimetre).
The surface condition of most measuring objects is
hardly ever perfectly even and homogeneous, even
though it might appear differently to the naked eye.
Observed under the microscope, even the smoothest
surface looks much like a mountain and valley
landscape. The teensiest of scratches, cavities or
contaminations can therefore already have a negative
effect on the expected measurement result, seeing as
they will also be a part of the measurement to a
greater or lesser extent. It does, however, not affect
the accuracy of the device. Even after the calibration
unexpected measurement deviations of a few
micrometres always have to be regarded in this
context.
Hence it is important to handle the supplied calibration
accessories with care to avoid scratches and dirt on
their respective surface as far as possible.
Zero point calibration
1. Press the Up button(2) for approx. 3s.
ðThe following message will be displayed:
Zero Reference
Place Probe
2. Position the sensor(4) on an uncoated spot of the material
to be measured. The measured medium is to be identical
to the coated material to be measured thereafter in terms
of material composition, shape and surface condition.
ðThe measurement will be acknowledged with an
acoustic signal.
ðThe measured value will be indicated on the display.
ðThe following message will be displayed:
Zero Reference
Lift off Probe
ðAnother acoustic signal will be emitted, then the
calibration is completed.
Carrying out a measurement
Carry out a zero point calibration before each measuring
operation.
1. If you want to save your measured values, use the Up and
Down buttons(2) to select a group (Group(1) to
Group(50)) or else to perform individual measurements
select Group(0).
2. Select the desired measuring mode (see Setting the
measuring mode).
3. Position the sensor(4) on the material to be measured and
perform group measurements or individual measurements.
ðThe measured value will be indicated on the display(1).
ðThe measurement will be acknowledged with an
acoustic signal.
Adjusting the storage settings for the measured values
The measured values are stored in groups.
The measured values can be saved in the groups from
1 to 50.
If you have selected Group(0), individually measured
values will only be displayed, but not saved.
In each group up to 50 measured values can be stored.
The groups can be selected in the start screen or else via
the Working Mode submenu.
The saved measured values can be deleted individually or
in groups.
Selecting a group:
1. Select the desired group in the start screen by use of the
Up or Down button(2).
2. As alternative, press the Left button(2) to call up the main
menu.
3. Use the Up or Down button(2) to select the menu item
Working Mode.
4. Confirm your selection by pressing the Left button(2).
ðA listing of the groups will be displayed.
5. Select the desired group by use of the Up or Down
button(2).
6. Confirm your selection by pressing the Left button(2).
ðThe main menu will be displayed.
Deleting measured values:
1. Press the Left button(2) to call up the main menu.
2. Use the Up or Down button(2) to select the menu item
Measure View.
3. Confirm your selection by pressing the Left button(2).
ðThe submenu Measure View will be displayed.
4. If you want to delete all measured values, select the menu
item Delete All, otherwise use the Up or Down button(2) to
select a group if you want to delete the measured values of
a certain group.
7


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Trotec BB30 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Trotec BB30 in the language / languages: English as an attachment in your email.

The manual is 1,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Trotec BB30

Trotec BB30 User Manual - German - 18 pages

Trotec BB30 User Manual - Dutch - 18 pages

Trotec BB30 User Manual - French - 18 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info