layer thickness measuring device BB20
General information on measurements
If the calibration has been carried out correctly, the measured
value lies within the guaranteed measuring tolerance. Incorrect
measured values can be deleted in the menu. The last value
results from the statistical calculation and the guaranteed
degrees of tolerance of the measured value.
Strong electrical or electromagnetic fields can influence the
measured values.
The measuring tip is to be cleared of contamination prior to
each calibration.
Wait for at least 4seconds after each measurement before
proceeding to the next measurement as the device does not
respond to measurement sequences carried out too quickly.
Deleting functions
In order to delete measured values, select Delete in the menu,
then select the measured values you would like to delete.
Current Data
Deletes the last measured value.
All Data
All data can be deleted during the respective working
mode.
Group Data
Comprises Delete all data. In addition, the limit values and
one-point and two-point calibration values will be deleted.
Limit value functions
Limit values can be entered using the Limit function. It is
possible to enter limit values before, during or after carrying out
measurement series.
1. Press the Red button(9) to open the main menu.
2. Select the Limit menu item, then select Limit setting.
ðUsing the (3) and (8) buttons, you can
determine the upper limit value in the High limit section
and the lower limit value in the Low limit section.
Each measured value beyond the fixed tolerance limit appears
on the display as follows:
H: Measured value is above upper limit value.
L: Measured value is below lower limit value.
Statistical measurements
The device is able to calculate statistics based on
80measurements. A total of 400measured values can be
saved.
No measured values are saved in direct mode, however,
calculating statistics is possible.
If you change between the different working modes or if the
device is switched off, statistics are deleted from the direct
mode.
The following statistical values are calculated:
NO.: Number of measured values in working mode
AVG: Average
Sdev.: Standard deviation
MAX: Maximum value
MIN: Minimum value
Memory space full
If the memory space is full in group mode, statistics will no
longer be updated. However, measurements can still be carried
out. These measurements will not be part of the statistics.
In single measurement mode, FULL appears on the display.
If the memory space is full in direct mode, old measurements
are replaced by new ones and statistics will be updated.
9


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Trotec BB20 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Trotec BB20 in the language / languages: English as an attachment in your email.

The manual is 0,55 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Trotec BB20

Trotec BB20 User Manual - German - 13 pages

Trotec BB20 User Manual - Dutch - 11 pages

Trotec BB20 User Manual - French - 11 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info