8 EN
layer thickness measuring device BB20
Zero point calibration
üAn uncoated sample is required.
1. Place the sensor onto an uncoated sample. The calibration
sample is to be identical with the actual sample in terms of
material composition, shape and surface condition.
ðThe following message will be displayed:
<x.x μm>
2. Lift the sensor again and keep it away from the sample (at
least 10cm).
3. Press and hold the Zero button(4) for approx. 2seconds.
ðThe following result appears on the display:
<0.0μm>
CAL1
ZeroY
ðThe calibration is terminated.
One-point calibration
This calibration is suited for high-precision measurements, e.g.
for thin coatings.
üA calibration film, a coated and an uncoated sample are
required.
1. Activate the calibration mode as described before.
2. Carry out a zero-point calibration.
3. Place the calibration film on an uncoated sample.
4. Place the sensor on the sample.
ðThe measurement is effected.
5. Lift the sensor again.
6. Adjust the required film thickness by using the Down/
Right(3) or the Up/Left(8) button. The film thickness
should be at least the size of the estimated coating
thickness.
7. Repeat this step several times.
8. Now carry out the measurement on a coated sample.
9. Press the Zero button(4) to adopt the current calibration.
10. To exit the current calibration, press the Blue button(2).
Two-point calibration
This calibration is especially suited for measurements on rough
surfaces or for high-precision measurements.
üYou need two films of different film thickness. The thicker
film should be approx.1.5 times thicker than the thinner
film.
1. At first, carry out a zero-point calibration.
2. Using the first film, carry out a one-point calibration.
3. Repeat this step with the second film.
4. Place the sensor onto the coating that is to be measured.
After a brief acoustic signal is emitted, you can lift the
sensor again.
ðThe measured value appears on the display.
Calibration on radiated surfaces
Values of layer thickness measurements carried out on radiated
surfaces are usually too high. The average thickness can be
determined as follows:
Method A
1. Carry out the steps for one-point and two-point calibration
as described above. Use a smooth sample with the same
surface curvature and material type as the measuring
object that is to be used later on.
2. In order to determine the average value Xo, carry out
approx.10 measurements on the uncoated sample.
3. Afterwards, carry out approx.10 measurements on the
coated sample to determine the average value Xm.
ðThe difference between both average values is the
average coating thickness Xeff. Take the standard
deviation of both values Xm and Xo into account.
Xeff= (Xm - Xo)±S
Method B
1. Carry out a zero calibration with approx.10measurements
on an uncoated sample.
2. Afterwards, carry out a film calibration on the uncoated
sample.
3. Repeat this calibration with films of different thicknesses
(max. thickness50μm). Together, these films should be as
thick as the assumed layer thickness.
ðThe layer thickness can be read from the display.
Method C
1. Carry out the steps for two-point calibration with two films
as described above.
2. Use several 50μm films to approach the actual surface as
precisely as possible.
ðThe layer thickness can be read from the display.
8


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Trotec BB20 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Trotec BB20 in the language / languages: English as an attachment in your email.

The manual is 0,55 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Trotec BB20

Trotec BB20 User Manual - German - 13 pages

Trotec BB20 User Manual - Dutch - 11 pages

Trotec BB20 User Manual - French - 11 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info