80-0067-00-MO-RevJ 2019-01-30 SunTech CT40 User Manual | 27
The sensor is too tight.
There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or sunlight.
A blood pressure cuff is inflated on the same extremity as the one with a SpO2 sensor attached.
The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia.
There is arterial occlusion proximal to the sensor.
The patient is in cardiac arrest or is in shock.
NOTE: See SP02 Sensor Specifications Table on page 45, the range of the peak wavelengths and maximum optical
output power of the light emitted by the pulse oximeter probe, which can be especially useful to clinicians.
NOTE: This device does not include an alarm system that detects an SpO2 or pulse rate physiological alarm condition.
This device is not designed for long-term monitoring and should only be used for spot-check measurements.
System Description:
The SunTech CT40 spot check device with pulse oximetry module consists of a base CT40 device for the display of
SpO2 measurement data and other user information, either a ChipOx or Masimo SET SpO2 module which is attached
to the CT40 base unit as described in Section 2, and the specific reusable pulse oximeter probe and/or cable that
comes packaged with each module. Device setup is described in more detail in Section 2. Both the Masimo® SET and
ChipOx SpO2 modules are calibrated to display functional oxygen saturation (SpO2).
Taking SpO2 Measurements:
1. Insert a finger (preferably the index, middle or ring finger) into the SpO2 sensor until the end of the finger reaches
the finger stop. Do not use the thumb.
2. The fingernail should face the side with the red light. Make sure that long fingernails do not interfere with proper
finger position
3. An SpO2 reading will be displayed after a few seconds. Remove the finger from the SpO2 sensor, and the last
measurement will be displayed and flashed.
4. If the sensor is not removed from the finger, SpO2 readings will be automatically stopped after 10 minutes of
continuous measurement, and the last measurement will be displayed and flashed.
5. To start a new measurement, re-apply the SpO2 sensor to the finger as described in Step 1.
NOTE: If the signal quality from the sensor drops below 90%, no measurement will be displayed in the SpO2 section
, and the percent symbol will begin flashing. Check sensor for correct placement and usage.
Special Notes for Masimo® SET SpO2 module:
NOTE: Possession or purchase of this device does not convey any express or implied license to use the device with
unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of one or
more of the patents relating to this device.
NOTE: If the accuracy of any measurement does not seem reasonable, first check the patient’s vital signs by alternate
means and the check the MS board pulse oximeter for proper functioning.
NOTE: A functional tester cannot be used to assess the accuracy of a Masimo pulse oximeter sensor or pulse oximeter.
Using Clinical Dynamics SmartSat SpO2 simulator, the device can display the entire range of SpO2 values.
Masimo® Sensors:
Before use, carefully read the M-LNCSTM DCI/DCIP sensor directions for use, which includes cleaning and disinfection
instructions.
27


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for SunTech CT40 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of SunTech CT40 in the language / languages: English as an attachment in your email.

The manual is 2,09 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of SunTech CT40

SunTech CT40 User Manual - German - 49 pages

SunTech CT40 User Manual - Dutch - 52 pages

SunTech CT40 User Manual - French - 50 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info