• Visual alarms remain open, but audible alarms are turned off.
• “Acknowledged” will appear in front of the acknowledged physiologic alarm message.
• Other remaining physiologic and technical alarms will remain active.
If a new technical or physiologic alarm occurs, the acknowledged alarms will not be influenced and the system will
produce an audible alarm according to the level of the new alarm condition.
6.9 Alarm Reset
Press the button on the front panel of the monitor to reset all active physiologic and technical alarms. When
alarms are reset, the following occurs:
• All auditory alarms are turned off.
• The visual alarm signals for any existing alarm conditions will continue as long as those alarm conditions
exist.
• Any technical alarms related to lead-off/sensor-off will be deleted.
• After alarms are reset, if a new technical alarm or physiologic alarm occurs, the monitor will enable the
audible alarm once again.
6.10 Alarm Audio Control
The following steps can turn the alarm audio function on or off.
Select 【SETTINGS】→【ADVANCED】, input the correct password to enter the alarm control interface. In this
interface, select 【Allow control alarm audio】. Return to the main interface and select【ALARM】 to choose 【
Alarm audio on 】or 【Alarm audio off】.
Note: After selecting
【
Alarm audio off
】
, the following icon will appear:
6.11 Reminder Signal
When alarm audio is turned off, the alarm system may be set to provide a periodic audible reminder signal sound
consisting of three chimes. Enter 【SETTINGS】→【ADVANCED】and input the correct password to enter the alarm
control interface. In this interface, select or deselect【Active reminder signal】to turn on or off the reminder signal.
The intervals between the reminder signals may be adjusted to be 30s, 60s, 90s, or 120s in this interface.
6.12 Resetting Alarm Limit
To reset all alarm limits to factory default levels, select 【Alarm】→【General】→【Reset alarm limits】. Limits
will be reset to the following defaults: