626431
14
Zoom out
Zoom in
Previous page
1/20
Next page
12
Guidance and Manufacturer’s Declaration – Immunity
The SurfLink Programmer is intended for use in the electromagnetic environment specified
below. The customer or user of the SurfLink Programmer should ensure that it is used in
such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic
Environment – Guidance
ESD
IEC 61000-4-2
±6 kV Contact
±8 kV Air
+/- 6 kV Contact
+/- 8 kV Air
+/- 6 kV Contact
+/- 8 kV Air
Note: Reset may
occur at 6 kV
Floors should be wood,
concrete or ceramic tile.
If floors are synthetic, the r/h
should be at least 30%.
EFT
IEC 61000-4-4
±2 kV Mains
±1 kV I/Os
Test not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV Differential
±2 kV Common
Test not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage
Dips/Dropout
IEC 61000-4-11
>95% Dip for
0.5 Cycle
60% Dip for
5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Test not
applicable
Mains power quality should
be that of a typical commercial
or hospital environment. If the
user of the SurfLink
Programmer requires
continued operation during
power mains interruptions,
it is recommended that the
SurfLink Programmer be
powered from an
uninterruptible power
supply or battery.
Power Frequency
50/60 Hz
Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic
fields should be that of a
typical
commercial or
hospital environment.
14


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Starkey SurfLink Programmer at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Starkey SurfLink Programmer in the language / languages: English as an attachment in your email.

The manual is 0,51 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info