Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV cont act
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
power supply lines:
±2 kV
input/output lines:
±1 kV
power supply lines:
±2 kV
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC61000-4-5
line(s) to line(s):
±1 kV
line(s) to earth:
±2 kV
100 kHz repetition
frequency
line(s) to line(s):
±1 kV
100 kHz repetition
frequency
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0% 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315°
0% 1 cycle
and 70% 25/30 cycles
Single phase: at 0
0% 300 cycle
0% 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315°
0% 1 cycle
and 70% 25/30 cycles
Single phase: at 0
0% 300 cycle
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Power frequency magnetic
fields should be at levels charac-
teristic of a typical location in a
typical commercial or hospital
environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
EMC GUIDANCE
167


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Soehnle 68096 - Systo Monitor Connect 300 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Soehnle 68096 - Systo Monitor Connect 300 in the language / languages: English, German, Dutch, Italian, Polish, Swedish, Spanish as an attachment in your email.

The manual is 1,56 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info