681498
19
Zoom out
Zoom in
Previous page
1/64
Next page
19
6. Operation
Caution
For hygiene reasons, the atomiser [9] and the ac
-
cessories must be cleaned after each treatment and
disinfected after the last treatment of the day.
The accessories may be used only by one person;
use by several people is not recommended.
If several different medicines are to be inhaled in
succession during therapeutic use, the atomiser [9]
must be rinsed under warm tap water after each
use. For information about this, see “Cleaning and
disinfection” on page 20.
Follow the instructions for changing the filter con-
tained in these instructions.
Before each use, check that the hose connections
on the nebuliser [6] and the atomiser [9] are tight.
Before use, check that the device is functioning cor-
rectly. To do this, switch on the nebuliser (including
the connected atomiser, but without any medication)
for a short time. If the atomiser [9] expels air, the
device is functioning correctly.
1. Inserting the nozzle attachment
To open the atomiser [9], twist the upper section
anti-clockwise against the medicine container [11].
Place the nozzle attachment [10] in the medicine
container [11].
Ensure that the cone for administering medication
fits well on the cone for the air duct inside the nebu-
liser.
2. Filling the atomiser
Pour an isotonic saline solu-
tion or the medicine directly
into the medicine container
[11]. Do not overfill! The max-
imum recommended fill level
is 8 ml !
Use medication only on the
advice of a physician, and
check the appropriate inhala-
tion duration and quantity for
your needs.
If the required volume of medication is less than 2 ml,
top up this quantity to at least 4 ml using an isotonic
saline solution. Viscous medicines may also require
dilution. Always follow the instructions of your doc
-
tor.
3. Closing the atomiser
To close the atomiser [9], twist the upper section
clockwise against the medicine container [11]. En-
sure the parts are connected correctly.
4. Connecting the accessories to the atomiser
C
onnect the required accessory part (mouthpiece
[12], adult mask [13] or child mask [14]) to the atom-
iser [9].
Note
The most effective form of nebulisation is by using the
mouthpiece. Nebulisation using a mask is only recom-
mended if it is not possible to use a mouthpiece (e.g.
for children who are not yet able to inhale medication
using a mouthpiece).
Before commencing treatment, pull the atomiser
upwards out of its holder [7].
Start the nebuliser using the on/off switch [3].
A flow of spray from the atomiser indicates that the
device is functioning properly.
5. Treatment
When inhaling, sit upright and relaxed at a table and
not in an armchair to avoid compressing the airways
and therefore impairing the effectiveness of the treat-
ment.
Inhale the nebulised medicine deeply.
Important
The device is not suitable for continuous operation;
after 30 minutes of operation it must be switched off
for 30 minutes.
Note
During treatment, hold the atomiser straight (vertical),
otherwise the atomisation does not work and correct
function cannot be guaranteed.
Important
Essential oils, cough medicines, solutions designed
for gargling, and drops for application to the skin or
for use in steam baths are wholly unsuitable for inha-
lation using a nebuliser. These substances are often
highly viscous and can impair the correct function of
the device and hence affect the effectiveness of its
application in the long term.
In the case of oversensitivity of the bronchial system,
medicines containing essential oils can sometimes
cause acute bronchiospasm (a sudden, cramp-like
tightening of the bronchi accompanied by breathless-
ness). Always ask your doctor or pharmacist for advice!
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Sanitas SIH 21 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Sanitas SIH 21 in the language / languages: English, German, French, Italian as an attachment in your email.

The manual is 1,73 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info