PRODUCT DESCRIPTION
Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse rate can be
measured automatically and non-invasively. The LCD display will show blood pressure and pulse rate. The most recent
2x60 measurements can be stored in the memory with date and time stamp. The monitor can also show the average
reading of the last three measurements.The Electronic Sphygmomanometers corresponds to the below standards: IEC
60601-1:2005/EN 60601-1:2006/AC:2010 (Medical electrical equipment -- Part 1: General requirements for basic safety
and essential performance), IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment -- Part 1-2:
General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -
Requirements and tests), IEC 80601-2-30: 2009+Cor.2010/EN 80601-2-30:2010(Medical electrical equipment –Part 2-30:
Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers),
EN 1060-1: 1995 + A1: 2002 + A2: 2009 (Non-invasive sphygmomanometers - Part 1: General requirements), EN 1060-3:
1997 + A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-
mechanical blood pressure measuring systems).
SPECIFICATIONS
Product name: Blood Pressure Monitor
Model BPA-9200-GB
Classification
Internally powered, Type BF applied part, IPX0, No AP or APG,
Continuous operation
Machine size Approx. 138mm × 98mm × 48mm (5 7/16”x 3 27/32”x 1 7/8”)
Cu circumference
30cm-42cm (11 13/16”-16 17/32”), 22cm-30cm (8 21/32”-11
13/16”) (Optional), 42cm-48cm (16 17/32”-18 29/32”)
(Optional)
Weight Approx. Approx. 211g (7 7/16 oz.) (exclude batteries and cu)
Measuring method Oscillometric method, automatic inflation and measurement
Memory volume 2x60 times with time and date stamp
Power source
Batteries: 4 ×1.5V Size AA
Measurement range
Cu pressure: 0-300mmHg
Systolic: 60-260mmHg
Diastolic: 40-199mmHg
Pulse rate: 40-180 beats/minute
Accuracy
Pressure: ±3mmHg
Pulse rate: ±5%
Environmental temperature for operation 10ºC~40ºC (50ºF~104ºF)
Environmental humidity for operation ≤90%RH
Environmental temperature for storage and transport -20ºC~55ºC (-4ºC~131ºC)
Environmental humidity for storage and transport ≤90%RH
Environmental pressure 80kPa-105kPa
Battery life Approx 500 times
All components belonging to the pressure measuring
system, including accessories
Pump, Valve, LCD, Cu, Sensor
Note: These specifications are subject to change without notice.
4


Need help? Post your question in this forum.

Forumrules
1

Forum

salter-bpa-9200

Reset search

  • is it possible to replace the salter monitor cuff model BPA-9200-GB and how much would that cost ?

    I keep getting Er 4 whenever I try to use it.. I have used it looser before without being told the problem is that I do not tighten it enough on my arm . Any advice please?
    Thank you Abbas Baba Submitted on 27-4-2021 at 16:45

    Reply Report abuse


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Salter BPA-9200 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Salter BPA-9200 in the language / languages: English as an attachment in your email.

The manual is 1,32 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info