Displaydescription problemWhere to look on MAIN PCBstatus OK
" " (nothing) = PCB stanbymode
0Switch armrest is in OFF positionX2b pin1, 2NC
1Switch armrest is in ON position (main safety-line is OK)
2Swivelseat-switch is operated (3 options in case of powered rotating seat)X2c pin 1,2NC
3Extracontact is not connected (via jumpers)X7 pin 3,4 and 5,6NC
4Current x10-3 is higher then 100 mAX10-3 (should not be connected) (when displayed at ONE tube lift exchange PCB)
5Lift is controlled on 2 placesX2a pin1,3 or X3a pin 2,3 or X2a pin 3,4 or X3a pin 5,6all NO
6Lift did not charge for 1 or 8 hours depending of dipswitch 3X10 pin 1,2 (pin 1 = voltage, pin 2 is (chassis)ground)20 ~30Vdc
7Lift came out of sleepmode / batterie connected and not controlled yet
ASafety-drive unit right is operatedX4 pin 1,2NC
bSafety-drive unit left is operatedX4 pin 3,4NC
CSafety-footrest right is operatedX5 pin 1,2NC
dSafety-footrest left is operatedX5 pin 3,4NC
ESafety-frame right is operated (normally jumpered in 1-tube lift)X6 pin 1,2NC
FSafety-frame unit left is operated (normally jumpered in 1-tube lift)X6 pin 3,4NC
gsafety endswitch is operated while lift is on opposite endpointX3a pin 1,2 at left endpoint or X3a pin 3,4 at right endpointNC
1
1
Safety endswitch right is operatedX3a pin 1,2NC
11
Safety endswitch left is operatedX3a pin 3,4
HHigh-load "overcurrent mainmotor" blokking/friction or simply too much weight
LLow batterie-voltageBATT + and BATT -22 <=> 29V
tTemperature PCB to high
PRotation monitor unit does not function correctl
y
X9a
p
in 1,2 IR-LED, X9a
p
in 3,4 switc
h
volta
g
e
p
in 3,4 alternatin
g
when movemen
t
y
p,,p,
gp,g
E01 too high voltage on X10-1. (wiringfault transformer/chargepoint)X10 pin 1,2 (pin 1 = voltage, pin 2 is (chassis)ground)
E02overvoltage batteriesexchange PCB (charging circuitry of measurement defect)
E03relay-errorexchange PCB
E04 other voltages in safety-line Main PCB
Connection of voltages to X3b, X9b, X2b, X2c or X7
E05interconnection between safetyline left and rightConnection between X3,X4,X5,X6 pin 1 or 2 to pin 3or 4
E06MAIN PCB defect (safetyline left or right are active when not controlled)exchange PCB
E07MAIN PCB defect (safetyline left or right will not active when controlled)exchange PCB
E08SMD fuse, contact catch (osg) or limit-switch not active
X9b 1 led off, X3b 2 led off (jumper), fuse or conection Bypass 3 LED off
E09Contacts on X7 do not work synchronously.X7 pin 3,4 and 5,6 (should be both jumpered)
E11Bypass is connected and activated by hold to run.
E12Batteries will not reach max voltage after 21 hours of charging
E13 *10 times overcurrent on output x10-3
E13 is stored directly (after 1 time overcuurent x10-3)
E14Internal defect PCB. exchange PCB
E15
Relay will not become active
(resistance in safetyline, Tiltsensor was active, relay defect ?)
E16Wiring brake Main motor not connected (when no fault reported by SCB)Check fault SCB (when NO fault => X1 brake NOT connected? Or ...Critical switch safetyline / Replace Main Board )
E17
Relay will not become active
(resistance in safetyline, Tiltsensor was active, relay defect ?)
E18 *Motorposition faultWhen seen multiple times or when this error prevents movement exchange PCB
E20Wrong dipswitch setting Main PCBDipswitch 7 + 8 on, and jumper ON,
E97Lift stopt by Sensorcontrol PCB (Check display on Sensorcontrol PCB) no beep
E98Error set by Sensorcontrol PCB (Check display on Sensorcontrol PCB) fast beep
[ NOTE * = no fast beep ]
E99 *No responce of sensorcontrol PCB Sensor control board in off state
Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.
Product:
Forumrules
To achieve meaningful questions, we apply the following rules:
First, read the manual;
Check if your question has been asked previously;
Try to ask your question as clearly as possible;
Did you already try to solve the problem? Please mention this;
Is your problem solved by a visitor then let him/her know in this forum;
To give a response to a question or answer, do not use this form but click on the button 'reply to this question';
Your question will be posted here and emailed to our subscribers. Therefore, avoid filling in personal details.
Register
Register getting emails for Otolift One at:
new questions and answers
new manuals
You will receive an email to register for one or both of the options.
Get your user manual by e-mail
Enter your email address to receive the manual of Otolift One in the language / languages: English as an attachment in your email.
The manual is 4,36 mb in size.
You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.
The manual is sent by email. Check your email
If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.
The email address you have provided is not correct.
Please check the email address and correct it.
Your question is posted on this page
Would you like to receive an email when new answers and questions are posted? Please enter your email address.