504816
9
Zoom out
Zoom in
Previous page
1/33
Next page
1716
How to disinfect the unit each day
Note: if the device is not cleaned and disinfected correctly and as frequently as indicated, micro-
organisms may remain in the unit, causing risk of infection.
The medication container, mesh cap, mouthpiece, mask and mask adapter should be disinfec-
ted
• before using the unit for the first time,
• after the unit has not been used for a long period of time, or
• daily during normal use.
Before disinfection, always first clean the parts as described in the section How to clean the unit
after each useon the previous page.
Disinfection of the individual parts can be done, using one of the following methods:
1. Disinfection by boiling.
Note: Main unit, medication container and flexible mask can NOT be boiled. For these parts, use
other methods, as described below.
Mesh-cap, mouthpiece and mask-adapter can be disinfected by boiling in distilled water. After 10
to 30 minutes of boiling, shake off excessive water from the parts and let them air dry on a clean,
lint free (paper) towel.
2. Disinfection by using alcohol.
All parts can be disinfected by using alcohol.
Note: Alcohol is highly flammable; do not use alcohol in close vicinity of open fire; do not smoke.
Soak the parts (NOT the main unit) in 70% ethyl alcohol for approx. 1 minute. Shake off exces
sive alcohol from the parts and let them air dry on a clean, lint free (paper) towel.
Alternatively, disinfect all parts by wiping them off with a lint-free cloth, soaked in alcohol.
3. Disinfection by commercially available disinfectant.
Note: Do NOT soak the main unit in any disinfectant.
Do not use disinfectant containing Sodium (Natrium) Hypochlorite; this may affect the metal parts.
For disinfection of the parts, refer to the instructions as supplied by the manufacturer of the disin-
fectant. After disinfection, thoroughly rinse the parts with warm tap water. Shake off excessive
water from the parts and let them air dry on a clean, lint free (paper) towel.
How to clean the unit after each inhalation
5. Wash the parts (not the main unit!) in a warm soapy solution.
Rinse the parts in warm tap water to remove the soap residue.
Some types of medications can easily remain on the mesh, therefore pay
close attention during washing.
Do not touch the mesh with a cotton swab, or with a pin.
6. Shake off excessive water from the parts and let them air dry on a
clean, lint free (paper) towel.
7. Clean the main unit and its electrodes by wiping it off with a (moistened) gauze.
Use a dry gauze to wipe the main unit dry.
Do not use volatile agents, such as benzene or thinner.
8. Assemble the unit, refer to page 10.
9. Store the unit in a dry, clean place.
9


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron MicroAIR - U22 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron MicroAIR - U22 in the language / languages: English, German, Dutch as an attachment in your email.

The manual is 1,55 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info