If a trouble occurs while using the unit, please check the following.
Problem What to inspect How to correct
The unit in ates to abnormally
high (low) pressure.
Is the arm cuff wrapped correctly?
Wrap the arm cuff correctly, and measure
again.
Is the patient moving his/her arm or body
during in ation?
Ask the patient not to move his/her arm or
body during measurement, and measure
again.
Does this trouble occur with a speci c
patient?
For patients having arrhythmia, the unit
may not in ate properly
Check the measurement by the stetho-
scope.
The unit cannot measure
blood pressure.
The blood pressure values are
extremely high (low).
First of all, check the patients condition.
After checking the patient with the stetho-
scope as needed, take a proper mea-
sure in accordance with the “list of error
codes”.
Is the patient moving his/her arm or body
during measurement?
Ask the patient not to move his/her arm or
body during measurement, and measure
again.
Does the patient have an arrhythmia? Check the patient with a stethoscope.
Is the size of the arm cuff and the wrapping
of arm cuff correct?
Select the arm cuff in accordance to the
circumference of the patients arm, wrap it
correctly, then measure again.
Is the level of the brachium to which the
arm cuff is wrapped at the same level as
the heart?
Keep the level of the brachium to which
the arm cuff is wrapped at the same level
as the heart, then measure again.
Troubleshooting
23


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron HEM-907 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron HEM-907 in the language / languages: English as an attachment in your email.

The manual is 3,28 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron HEM-907

Omron HEM-907 User Manual - German - 26 pages

Omron HEM-907 User Manual - Dutch - 26 pages

Omron HEM-907 User Manual - Dutch - 23 pages

Omron HEM-907 User Manual - Italian - 26 pages

Omron HEM-907 User Manual - Spanish - 26 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info