List of Error Codes
Error code Cause How to correct
Er1
In ation error
When the pressure does not exceed 15
mmHg after operating a pump for 15 seconds
from the start of in ation
When the in ation does not reach the set
pressure of the arm cuff within the speci ed
time after starting the in ation
If the connector of rubber tube connecting the
arm cuff and the main unit is loose, connect them
correctly, then measure.
If the rubber tube connecting the arm cuff and the
main unit is bent, stretch it, then measure.
If the arm cuff is wrapped loosely, wrap it correctly,
then measure.
Er2
De ation error
When the de ation speed is too fast during
the measurement
When the de ation speed is too slow during
the measurement
When the measurement does not end
within the speci ed time after starting the
measurement
When the pressure is applied more than 5
minutes
Er3
Overpressure error
The arm cuff pressure exceeded 299 mmHg.
If the rubber tube connecting the arm cuff and the
main unit is bent, stretch it, then measure.
Er4
Insuf cient in ation error
Blood pressure could not be measured due
to insuf cient in ation after the measurement
ends.
If the measurement is made by setting P-SET to
AUTO”, ask the patient not move his/her arm or
body during the in ation.
When setting the P-SET to “AUTO”, turn the Volume
counterclockwise as far as it goes until you can hear
the click sound.
Set the P-SET to 30 to 40 mmHg higher than the
systolic pressure of the patient, then measure.
Er5
Indeterminable blood pressure error
Blood pressure could not be measured even
when the arm cuff pressure reached the
speci ed pressure.
If the arm cuff is wrapped loosely, wrap it correctly,
then measure.
Er6
Pulse wave small error
Pulse wave was too small.
If the arm cuff is wrapped loosely, wrap it correctly,
then measure.
Do not wrap the arm cuff over a thick cloth or by
rolling up the jacket or shirt.
Er7
Measurement error
Relationship between systolic and diastolic
pressures was abnormal.
Check if the patient has an arrhythmia.
Ask the patient not to move his/her arm or body
during the measurement.
Er8
Pulse rate error
Pule rate did not stay within the range of 30
to 199 beats/min.
Er9
Device error
There is a problem with the main unit of the
blood pressure monitor.
Contact the dealer where you purchased the unit or
the nearest OMRON dealer.
22


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron HEM-907 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron HEM-907 in the language / languages: English as an attachment in your email.

The manual is 3,28 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron HEM-907

Omron HEM-907 User Manual - German - 26 pages

Omron HEM-907 User Manual - Dutch - 26 pages

Omron HEM-907 User Manual - Dutch - 23 pages

Omron HEM-907 User Manual - Italian - 26 pages

Omron HEM-907 User Manual - Spanish - 26 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info