549536
33
Zoom out
Zoom in
Previous page
1/38
Next page
31
EN
Environmental Conditions
Non-Invasive Blood Pressure (NIBP)
* Comparison with auscultation method performed by a trained professional.
DIA determined by the auscultation method is “K5”.
Note: Specifications may be changed without prior notice.
This blood pressure monitor fulfils the requirements of the EC directive 93/42/EEC (Medical Device
Directive). It also conforms to the European standard EN 1060, Non-invasive Sphygmomanometers
Part 1: General Requirements and Part 3: Additional Requirements for Electromechanical Blood
Pressure Measuring Systems.
Operational temperature and
humidity
Temperature range: 10 to 40°C (50 to 104°F)
Humidity range:
30 to 85%RH (not condensed)
Atmospheric pressure: 700 to 1060hPa
Storage and transportation Temperature range: -20 to 60°C (-4 to 140°F)
Humidity range: 10 to 95%RH (not condensed)
Atmospheric pressure: 500 to 1060hPa
Measurement technology Oscillometric
Measurement method Dynamic Linear Deflation method
Pressure display range 0 to 300 mmHg
Pressure display accuracy Within ±3 mmHg
NIBP measurement range SYS 60 to 250 mmHg
DIA 40 to 200 mmHg
Pulse rate 40 to 200 /min
NIBP accuracy* Maximum mean error within ±5mmHg
Maximum standard deviation within 8mmHg
Pulse rate accuracy Within ±5 % of reading
Reference standard: EN1060-1:1995+A2:2009
EN1060-3:1997+A2:2009
ISO81060-1:2007
33


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron HBP-1300 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron HBP-1300 in the language / languages: English as an attachment in your email.

The manual is 0,53 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron HBP-1300

Omron HBP-1300 User Manual - German - 38 pages

Omron HBP-1300 User Manual - Dutch - 38 pages

Omron HBP-1300 User Manual - French - 38 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info