Non-Invasive Blood Pressure (NIBP) Measurement
Non-Invasive Pressure Measurement Principles
Oscillometric method
The beat in the pulsation generated by the contraction of the heart is captured as the pressure inside
the cuff to measure the blood pressure. If the cuff wrapped around the upper arm is pressurized
sufficiently, the blood flow stops, but the beat of the pulsation is present and the pressure inside the cuff
receives this and oscillates. Next, as the pressure inside the cuff gradually decreases, the oscillation of
the pressure within the cuff gradually increases and reaches a peak. As the pressure within the cuff
decreases further, the oscillation decreases from its peak.
The pressure within the cuff and the relationship with the increase and decrease of the oscillation within
the cuff in this series of processes are stored into memory, calculations are carried out, and the blood
pressure value is determined.
The pressure within the cuff when the oscillation increases drastically is the systolic pressure and the
pressure within the cuff when the oscillation decreases drastically is the diastolic pressure. Also, the
pressure within the cuff when the oscillation peaks is taken as the average pulsation pressure.
The oscillometric method does not determine the blood pressure value instantly like a microphone type
automatic blood pressure gauge with the auscultation method, but rather determines it from the series
of change curves as explained above. Therefore, it is not easily affected by external noise, an electric
scalpel or other electro surgical instruments.
KOROTKOV SOUNDS
RADIAL PULSE
5 SEC
OSCILLATIONS IN CUFF PRESSURE
CUFF
PRESSURE
Comparison between the auscultatory, oscillometric and
palpatory methods of measuring blood pressure.
L.A. Geddes,
The Direct and Indirect Measurement of Blood Pressure”, Year Book Medical Publishers, Inc. 1970
20


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Omron HBP-1300 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Omron HBP-1300 in the language / languages: English as an attachment in your email.

The manual is 0,53 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Omron HBP-1300

Omron HBP-1300 User Manual - German - 38 pages

Omron HBP-1300 User Manual - Dutch - 38 pages

Omron HBP-1300 User Manual - French - 38 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info