Avant 9600 Option switches
The Avant 9600 contains eight option switches that are located under the blue back
cover of the unit (See “Installing the Batteries” for instructions on removing the
back cover). The UP position faces the top of the unit, and the DOWN position
faces the bottom of the unit. The factory setting for all Avant 9600 option
switches is the DOWN position.
Switch Function
Switch 1 Alarm Disable Lock
Up—Alarm volume may be disabled
Down—Alarm volume cannot be set to zero
Switch 2 Normal / Slow SpO
2
Averaging
Up—Slow Averaging (8 beat exponential average)
Down—Normal Averaging (4 beat exponential average)
Switch 3 Alarm Unlatched / Latched
Up—Alarms Latched
Down—Alarms Unlatched
Switch 4 Factory / User-Defined Defaults
Up—User-Defined Defaults for Alarm Limits and Volume Settings
Down—Factory Defaults for Alarm Limits and Volume Settings
Switch 5
US / International Date Format
Up—International Date Format
Down—US Date Format
Switch 6 Power Saving Display Dim During Battery Operation
Up—Power Saving Disabled
Down—Power Saving Enabled
Switch 7
Nurse Call Output
Up—Continuous
Down—Momentary
Switch 8 Serial Patient Data Output
Up—Fast responding SpO
2
and pulse rate output
Down—SpO
2
and pulse rate output, as selected by option switch 2
12
3
4
5
6
78
19


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nonin Avant 9600 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nonin Avant 9600 in the language / languages: English as an attachment in your email.

The manual is 0,7 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

Others manual(s) of Nonin Avant 9600

Nonin Avant 9600 User Manual - German - 47 pages

Nonin Avant 9600 User Manual - Dutch - 25 pages


The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info