SECTION III OPERATION MANUAL
Utrecht
- 35 - Ref 002
METER Configuration
Irange(1V) : 1000A
Uinp.K.(*) : 1
sync.inpt. : auto
last calb.:
18.06.2001 09:18
connection : 4wire
sync.freq : 50 Hz
Unomin.(V) : 230.0 V
Fig. 12: Meter Configuration Sub-menu
Use the UP and DOWN keys to select the required parameter.
UNOMIN.(v) Range:
50.0 V to
450.0 V
Nominal measuring range of instrument voltage inputs. It is
used for calculation and display of results only.
Default value is 230.0 V.
Uinp.K.(*) Range:
1 to 800
Scaling factor for voltage inputs.
This allows for external voltage transformers or dividers that
may be used and ensures that readings are related to the
primary.
Example: for 11 kV / 110 V, the multiplication factor must be set
to 100.
Use the LEFT and RIGHT keys to set Uinp.K.
Standard and default value is 1.
Displayed full scale voltage range is UNOMIN. * UINP.K.
Irange (1V) Range:
1 A to 30 kA
Scale factor for current inputs.
Defines the current equivalent to a 1 V input signal.
Use the LEFT and RIGHT keys to set Irange.
Standard and default value is 1000 A.
connection Defines the method of connecting the Instrument to the 3ϕ
systems:
4 wire 3ϕ 4-wire system (with a Neutral conductor).
All voltage and current inputs are used.
3wire 3ϕ 3-wire system (without Neutral conductor) 3 current
transformers used.
AARON
3ϕ 3-wire system (without Neutral conductor) (also known as
the ‘2 wattmeter method’) 2 current transformers used.
Press SELECT key for connection type selection.
sync. freq. 50 Hz, 60 Hz
Default mains frequency for input cycle period/scanning.
It is ignored when the instrument detects valid frequency on the
selected sync. input.
Press SELECT key for selection of system frequency.
sync. inp. U1, U2, U3, I1,
AUTO
Default synchronisation input.
Use fixed input for synchronisation or auto detect mode
(automatic scanning for a valid synchronisation input).
36


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Nieaf-Smitt PQA 6600 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Nieaf-Smitt PQA 6600 in the language / languages: English as an attachment in your email.

The manual is 1.72 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info