5.
Press the START button
3
and steadily move the
thermometer
from the middle of the forehead to the
temple area (about 1 cm above the eyebrow). After
3 seconds a long beep will verify the completion of
measurement. If the temple area hasn’t been reached
before the long beep is heard, repeat the measurement as
described above, but move the thermometer a little faster.
6. Read the recorded temperature from the LCD display.
Measuring in Object Mode
1. Follow steps 1-3 above, then aim the thermometer at the
center of the object you want to measure with a distance
of no more than 5 cm. Press the START button 3. After
3 seconds a long beep will verify the completion of
measurement.
2. Read the recorded temperature from the LCD display.
NOTE:
Patients and thermometer should stay in steady-
state room condition for at least 30 minutes.
Don't take a measurement while or immediately after
nursing a baby.
Don't use the thermometer in high humidity environments.
Patients should not drink, eat, or exercise before/while
taking the measurement.
Don't move the measurement device from the
measuring area before hearing the termination beep.
Use an alcohol swab to carefully clean the probe and
wait for 15 minutes before taking a measurement on
another patient.
10 short beeps and a red LCD backlight alert the patient
that he/she may have a temperature equal to or higher
than 37.5 °C.
Always take the temperature in the same location, since
temperature readings may vary according to locations.
Doctors recommend rectal measurement for newborn
infants within the first 6 months, as all other measuring
methods might lead to ambiguous results. If using a non
contact thermometer on those infants, we recommend to
always verify the readings with a rectal measurement.
In the following situations it is recommended that three
temperatures are taken and the highest one taken as the
reading:
1. Children under three years of age with a compro-
mised immune system and for whom the presence or
absence of fever is critical.
2. When the user is learning how to use the thermometer
for the first time until he/she has familiarized himself/
herself with the instrument and obtains consistent
readings.
3. If the measurement is surprisingly low.
Readings from different measuring sites should not
be compared as the normal body temperature varies
by measuring site and time of day
, being highest in the
evening and lowest about one hour before waking up.
Normal body temperature ranges:
- Axillar: 34.7 - 37.7 °C / 94.5 - 99.1 °F
- Oral: 35.5 - 37.5 °C / 95.9 - 99.5 °F
- Rectal: 36.6 - 38.0 °C / 97.9 - 100.4 °F
- Microlife NC 120: 35.4 °C - 37.4 °C / 95.7 - 99.3 °F
7. Changing between Celsius and Fahrenheit
This thermometer can display temperature measurements in
either Fahrenheit or Celsius. To switch the display between
°C and °F, simply turn OFF the unit,
press and hold
the
START button
3
for 5 seconds; after 5 seconds, the current
measurement scale («
°C
» or «
°F
» icon) will be flashing on
the display
AO
. Change the measurement scale between °C
and °F by pressing the START button
3
. When the measure-
ment scale has been chosen, wait for 5 seconds and the unit
will automatically enter the «ready for measuring» mode.
5


Need help? Post your question in this forum.

Forumrules


Report abuse

Libble takes abuse of its services very seriously. We're committed to dealing with such abuse according to the laws in your country of residence. When you submit a report, we'll investigate it and take the appropriate action. We'll get back to you only if we require additional details or have more information to share.

Product:

For example, Anti-Semitic content, racist content, or material that could result in a violent physical act.

For example, a credit card number, a personal identification number, or an unlisted home address. Note that email addresses and full names are not considered private information.

Forumrules

To achieve meaningful questions, we apply the following rules:

Register

Register getting emails for Microlife NC 120 at:


You will receive an email to register for one or both of the options.


Get your user manual by e-mail

Enter your email address to receive the manual of Microlife NC 120 in the language / languages: English, German, Dutch, Danish, French, Italian, Swedish, Norwegian, Finnish as an attachment in your email.

The manual is 3,3 mb in size.

 

You will receive the manual in your email within minutes. If you have not received an email, then probably have entered the wrong email address or your mailbox is too full. In addition, it may be that your ISP may have a maximum size for emails to receive.

The manual is sent by email. Check your email

If you have not received an email with the manual within fifteen minutes, it may be that you have a entered a wrong email address or that your ISP has set a maximum size to receive email that is smaller than the size of the manual.

The email address you have provided is not correct.

Please check the email address and correct it.

Your question is posted on this page

Would you like to receive an email when new answers and questions are posted? Please enter your email address.



Info